Severity: Warning
Message: file_get_contents(https://...@pubfacts.com&api_key=b8daa3ad693db53b1410957c26c9a51b4908&a=1): Failed to open stream: HTTP request failed! HTTP/1.1 429 Too Many Requests
Filename: helpers/my_audit_helper.php
Line Number: 176
Backtrace:
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 176
Function: file_get_contents
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 250
Function: simplexml_load_file_from_url
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 1034
Function: getPubMedXML
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 3152
Function: GetPubMedArticleOutput_2016
File: /var/www/html/application/controllers/Detail.php
Line: 575
Function: pubMedSearch_Global
File: /var/www/html/application/controllers/Detail.php
Line: 489
Function: pubMedGetRelatedKeyword
File: /var/www/html/index.php
Line: 316
Function: require_once
Background: Telescopic rods require alignment of 2 rods to enable lengthening. A telescopic rod converts functionally into a solid rod if either rod bends, preventing proper engagement. Our goal was to characterize implant bending as a mode of failure of telescopic rods used in the treatment of osteogenesis imperfecta in children.
Methods: We conducted a retrospective review of our osteogenesis imperfecta database for patients treated with intramedullary telescopic rods at our institution from 1992 through 2010 and identified 12 patients with bent rods. The 6 boys and 6 girls had an average age at the time of initial surgery of 3.1 years (range, 1.8 to 8.3 y) and a total of 51 telescoping rods. Clinic notes, operative reports, and radiographs were reviewed. The rods were analyzed for amount of lengthening, characteristics of bending, presence of cut out, or disengagement from an anchor point. Bends in the rods were characterized by their location on the implant component. The bent and straight rods were compared. Data were analyzed with the Mann-Whitney test (statistical significance set at P≤0.05).
Results: Of the 51 telescoping rods, 17 constructs (33%) bent. The average interval between surgery and rod bending was 4.0 years (range, 0.9 to 8.2 y). Before bending, 11 of 17 telescoping rods had routine follow-up radiographs for review. In 10 of the rods, bending was present when early signs of rod failure were first detected. Rod bending did not seem to be related to rod size. There was no area on the rod itself that seemed more susceptible to bending.
Conclusions: Rod bending can be an early sign of impending rod failure. When rod bending is first noted, it may predispose the rod to other subsequent failures such as loss of proximal and distal fixation and cut out. Rod bending should be viewed as an indicator for closer monitoring of the patient and discussions regarding future need for rod exchange.
Level Of Evidence: Level III-retrospective review.