Objective: To assess foetal absorbed radiation dose (FAD) during prophylactic occlusion balloon catheter (POBC) placement before surgery for morbidly adherent placenta (MAP).

Methods: Retrospective cohort study of women with MAP undergoing POBC before surgery. Women were divided into two groups. Group 1 treated between 2008 and 2011, Group 2 between 2012 and 2014 when fluoroscopy pulse rate reduced from 7.5 to 2 pulses per second. Median and interquartile range (IQR) of dose area product (DAP) and FAD were compared using the Mann-U-Whitney test. FAD was calculated using the PCXMC Monte-Carlo calculation. All procedures were performed using the same equipment (Siemens Axiom Artis dTA) by the same experienced interventional radiology team.

Results: Thirty-four women underwent POBC. 16 procedures were performed in Group 1 and 18 in Group 2. Median DAP overall was 1025.10 µGym(2) (IQR 532.1-1775.3): Group 1, 1713.25 µGym(2) (IQR 1164.5-2274.5) and Group 2, 660.70 µGym(2) (IQR 440.9-1020.9) (p = 0.027). Median FAD overall was 4.65 mGy (IQR 2.7-8.2): Group 1, 6.25 mGy (IQR 4.4-10.6) and Group 2, 3.05 mGy (IQR 2.0-5.4) (p = 0.031).

Conclusions: FAD during POBC insertion is low overall. The only change between the two groups was in pulsed fluoroscopy rate which resulted in a significant reduction in FAD. This reinforces the importance of implementation of dose reduction techniques.

Download full-text PDF

Source
http://dx.doi.org/10.1007/s00270-015-1102-7DOI Listing

Publication Analysis

Top Keywords

µgym2 iqr
12
mgy iqr
12
radiation dose
8
prophylactic occlusion
8
occlusion balloon
8
morbidly adherent
8
adherent placenta
8
group
8
procedures performed
8
iqr
7

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!