A PHP Error was encountered

Severity: Warning

Message: file_get_contents(https://...@pubfacts.com&api_key=b8daa3ad693db53b1410957c26c9a51b4908&a=1): Failed to open stream: HTTP request failed! HTTP/1.1 429 Too Many Requests

Filename: helpers/my_audit_helper.php

Line Number: 176

Backtrace:

File: /var/www/html/application/helpers/my_audit_helper.php
Line: 176
Function: file_get_contents

File: /var/www/html/application/helpers/my_audit_helper.php
Line: 250
Function: simplexml_load_file_from_url

File: /var/www/html/application/helpers/my_audit_helper.php
Line: 3122
Function: getPubMedXML

File: /var/www/html/application/controllers/Detail.php
Line: 575
Function: pubMedSearch_Global

File: /var/www/html/application/controllers/Detail.php
Line: 489
Function: pubMedGetRelatedKeyword

File: /var/www/html/index.php
Line: 316
Function: require_once

A clinical and laboratory approach used to elucidate discordant results of high-sensitivity troponin T and troponin I. | LitMetric

A clinical and laboratory approach used to elucidate discordant results of high-sensitivity troponin T and troponin I.

Clin Chim Acta

Department of Laboratory Methods, Institute for Clinical and Experimental Medicine, Vídeňská 1958/9, 140 21 Prague 4, Czech Republic; Charles University, 3rd Faculty of Medicine, Ruská 87, 100 00 Prague 10, Czech Republic. Electronic address:

Published: June 2015

Background: Careful interpretation of discordant results in high-sensitivity troponin measurements is necessary in cases of suspect immunoassay interferences. We describe several procedures taken in a case of a polymorbid patient with chest pain, without clear evidence of myocardial necrosis and with increased high-sensitivity cardiac troponin T (hs-cTnT). We checked the Vafaie's algorithm for the evaluation of suspect interference in troponin measurements.

Methods: We conducted a case report analysis, additional measurements, a dilution test and pretreatment of plasma with blocking agents.

Results: Concentration of hs-cTnT (99 th percentile of "healthy" population 14 ng/L) increased from 120.1 ng/L to 280.4 ng/L during an 8-month period and decreased to 216.3 ng/L during the following month with repeatedly negative troponin I (TnI), hs-cTnI, myoglobin and creatine kinase MB (CK-MB). Suspected false positivity of hs-cTnT was further confirmed by treatment of plasma with an antiheterophile blocking agent (hs-cTnT before treatment 280.4 ng/L, after 16.53/16.23 ng/L). This outcome was further confirmed by the manufacturer's experiments.

Conclusions: The false-positive results of hs-cTnT were caused by the presence of extremely rare high molecular weight protein, presumably IgM, most likely HAMA (human anti-mouse antibody). Only the pre-treatment of plasma with a blocking agent provided a reliable indication of the interference. Cooperation among clinicians, laboratory personnel and the manufacturer is essential.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.cca.2015.03.046DOI Listing

Publication Analysis

Top Keywords

discordant high-sensitivity
8
high-sensitivity troponin
8
plasma blocking
8
2804 ng/l
8
blocking agent
8
troponin
6
ng/l
6
hs-ctnt
5
clinical laboratory
4
laboratory approach
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!