Purpose: The aim of this study was to reduce the exposed dose of radiotherapy treatment planning computed tomography (CT) by using low tube voltage technique.

Materials And Methods: We used tube voltages of 80 kV, 100 kV, and 120 kV, respectively. First, we evaluated exposure dose with CT dose index (CTDI) for each voltage. Second, we compared image quality indexes such as modulation transfer function (MTF), noise power spectrum (NPS), and contrast to noise ratio (CNR) of phantom images with each voltage. Third, CT to electron density tables were measured in three voltages and monitor unit value was calculated along with clinical cases. Finally, CT surface exposed dose of chest skin was measured by thermoluminescent dosimeter (TLD).

Results: In image evaluation MTF and NPS were approximately equal; CNR slightly decreased, 2.0% for 100 kV. We performed check radiation dose accuracy for each tube voltage with each model phantom. As a result, the difference of MU value was not accepted. Finally, compared with 120 kV, CTDIvol and TLD value showed markedly decreased radiation dose, 60% for 80 kV and 30% for 100 kV.

Conclusion: Using a technique with low tube voltages, especially 100 kV, is useful in radiotherapy treatment planning to obtain 20% dose reduction without compromising 120 kV image quality.

Download full-text PDF

Source
http://dx.doi.org/10.6009/jjrt.2015_JSRT_71.4.308DOI Listing

Publication Analysis

Top Keywords

treatment planning
12
low tube
12
tube voltage
12
dose
8
exposure dose
8
dose reduction
8
exposed dose
8
radiotherapy treatment
8
tube voltages
8
voltages 100
8

Similar Publications

Background: HIV/AIDS remains a significant global challenge, and with the rapid advancement of technology, there has been an increasing number of interventions aimed at improving HIV/AIDS cognition and self-management behaviors among patients. However, there is still a lack of detailed literature integrating relevant evidence.

Objective: This study aims to comprehensively review existing research on interventions using modern information methods to improve HIV/AIDS cognition and enhance self-management behaviors among patients.

View Article and Find Full Text PDF

Background: Establishing and maintaining gamma knife facility incurs significant costs, mandating healthcare institutions to meticulously assess financial implications for sustainability.

Methods: This study explores the financial implications of setting up and operating a Gamma Knife facility, with an aim to ascertain user charges for achieving breakeven. The study was conducted from January to June 2019 at the largest neurosurgery centre of an Institute of National Importance (INI), in Delhi, India.

View Article and Find Full Text PDF

Type D personality, characterized by negative affectivity and social inhibition, has been associated with both the psychophysiology of coronary artery disease (CAD) and depressive disorders. However, few reports have described the impact of coping strategies in these patients. This study aimed to analyze the characteristics of type D personality and the coping strategies adopted by patients with CAD and to explore the factors associated with depressive tendencies during follow-up.

View Article and Find Full Text PDF

Background: Minimizing the duration of mechanical ventilation is one of the most important therapeutic goals during the care of preterm infants at neonatal intensive care units (NICUs). The rate of extubation failure among preterm infants is between 16% and 40% worldwide. Numerous studies have been conducted on the assessment of extubation suitability, the optimal choice of respiratory support around extubation, and the effectiveness of medical interventions.

View Article and Find Full Text PDF

Forecasting the use of chiropractic services within the Veterans Health Administration.

PLoS One

January 2025

Biomedical Informatics and Data Science, Yale School of Medicine, New Haven, CT, United States of America.

Objective: To model future use of chiropractic services and predict clinical resource needs within the Veterans Health Administration (VA) over the next 5 years.

Methods: A serial cross-sectional analysis of chiropractic use data from VA's Corporate Data Warehouse for fiscal years (FY) 2017 through 2022 (10/1/2016-9/30/2022). We calculated the proportion of VA chiropractic users-via care provided on-station and/or purchased from Community Care Network (CCN) providers-compared to overall VA healthcare users for each FY.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!