Purpose: The purpose of the present study was to introduce and evaluate the use of a modified "wax-bite dental splint" for surgical navigation in craniomaxillofacial surgery.
Materials And Methods: A 2-layer wax bite dental splint was fabricated with an anterior extension, and 8 gutta percha markers were incorporated in each splint for marker-based pair-point registration. To evaluate the accuracy, the occlusal registration of the wax bite dental splint was performed on 10 artificial skulls. Consecutively, all the skulls were scanned using a standardized cone-beam computed tomography scanning protocol with the 2-layer wax bite dental splint in place. Automatic rigid marker-based pair-point registration was performed using the iPlan CMF software, version 3.0 (Brainlab AG, Feldkirchen, Germany) using 8 gutta percha markers incorporated into the wax bite dental splint. Additionally, the registration accuracy of 6 anatomic skeletal landmarks was measured on each skull.
Results: The mean registration error for each wax bite dental splint ranged from 0.78 to 1.01 mm. The overall mean registration error for the wax-bite dental splint-based registration was 0.89 ± 0.08 mm. The mean registration error for the 6 anatomic landmarks ranged from 1.23 to 2.3 mm. The overall mean registration error was 1.68 ± 0.28 mm.
Conclusion: The results of the present study show the potential for a wax-bite dental splint as an alternative rigid registration method for surgical navigation in craniomaxillofacial surgery. Moreover, from a clinical viewpoint, the method is accurate, user-friendly, inexpensive, and not time-consuming.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1016/j.joms.2015.03.057 | DOI Listing |
Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!