A PHP Error was encountered

Severity: Warning

Message: file_get_contents(https://...@pubfacts.com&api_key=b8daa3ad693db53b1410957c26c9a51b4908&a=1): Failed to open stream: HTTP request failed! HTTP/1.1 429 Too Many Requests

Filename: helpers/my_audit_helper.php

Line Number: 176

Backtrace:

File: /var/www/html/application/helpers/my_audit_helper.php
Line: 176
Function: file_get_contents

File: /var/www/html/application/helpers/my_audit_helper.php
Line: 250
Function: simplexml_load_file_from_url

File: /var/www/html/application/helpers/my_audit_helper.php
Line: 1034
Function: getPubMedXML

File: /var/www/html/application/helpers/my_audit_helper.php
Line: 3152
Function: GetPubMedArticleOutput_2016

File: /var/www/html/application/controllers/Detail.php
Line: 575
Function: pubMedSearch_Global

File: /var/www/html/application/controllers/Detail.php
Line: 489
Function: pubMedGetRelatedKeyword

File: /var/www/html/index.php
Line: 316
Function: require_once

Is 3D MPRAGE better than the combination DIR/PSIR for cortical lesion detection at 3T MRI? | LitMetric

Background And Objectives: Based on the application of newer magnetic resonance imaging (MRI) acquisition sequences, the detection of cortical lesions (CL) in multiple sclerosis (MS) has significantly improved. Double inversion recovery (DIR) at 3T has increased the detection sensitivity and classification specificity when combined with phase sensitive inversion recovery (PSIR). Previous findings with 3D magnetization prepared rapid acquisition with gradient echo (MPRAGE) sequences, showed improved classification specificity of purely intracortical (IC) and mixed (MX) lesions, compared to the classification based on DIR/PSIR. Direct comparison between the detection of CL by 3D MPRAGE and by DIR/PSIR at 3T has not been evaluated.

Methods: Eleven subjects were imaged on a 3T magnet. DIR/PSIR and 3D MPRAGE images were reviewed independently. Each image set was reviewed twice; only lesions detected on both sessions were scored. Review time per scan was ~5min for DIR/PSIR and ~15min for 3D MPRAGE.

Results: We identified 141 CL (62 IC+79 MX) based on DIR/PSIR images vs. 93 (38 IC+55 MX) based on MPRAGE from all eleven patients. MPRAGE under-detected the number of CL in seven cases and over-detected the number of CL in three, only one case had the same number of CL on both sets of images.

Conclusions: Combination DIR/PSIR at 3T is superior to 3D MPRAGE for detection of cortical gray matter lesions in MS. The contrast-to-noise ratio of CL appears to be inferior on the MPRAGE images relative to DIR/PSIR.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.msard.2013.10.002DOI Listing

Publication Analysis

Top Keywords

mprage
8
dir/psir
8
combination dir/psir
8
detection cortical
8
inversion recovery
8
classification specificity
8
based dir/psir
8
mprage images
8
detection
5
mprage better
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!