Purpose: To report an unusual late type Ia endoleak after endovascular aneurysm repair with a low-profile sealing ring endograft and its treatment with transcaval sac embolization.

Case Report: An 82-year-old man was treated for a 47-mm abdominal aortic aneurysm associated with a 35-mm left common iliac artery aneurysm using an Ovation Abdominal Stent Graft System. Correct endograft placement was documented at 3-month follow-up. Seven months after the repair, the patient was admitted to the emergency department with acute left inguinal pain. Computed tomography showed aneurysm sac enlargement (57 mm) and a contained rupture of the left common iliac artery due to a late type Ia endoleak. The endoleak was embolized with coils and thrombin via a transcaval approach. Sac shrinkage without signs of endoleak was observed at 12-month follow-up.

Conclusion: In this case, late endograft failure with sealing ring collapse created an inflow channel into the aneurysm sac, with massive type Ia endoleak.

Download full-text PDF

Source
http://dx.doi.org/10.1177/1526602815579898DOI Listing

Publication Analysis

Top Keywords

type endoleak
16
late type
12
treatment transcaval
8
sealing ring
8
left common
8
common iliac
8
iliac artery
8
aneurysm sac
8
endoleak
6
aneurysm
5

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!