Purpose: To assess the correlations between the relative afferent pupillary defect measured by a binocular, computerized, and commercially available pupillometer and the inter-eye differences in average retinal nerve fiber layer (RNFL) thickness measured by spectral-domain optical coherence tomography (SD-OCT) and visual field index (VFI) and mean deviation (MD) measured by automated static perimetry in patients with glaucoma.

Design: Cross-sectional study.

Participants: Thirty four patients with open-angle glaucoma.

Methods: Pupillary constriction amplitudes were measured with an automated pupillometer. SD-OCT imaging of the circumpapillary RNFL and automated static perimetry were performed within 6 months of pupillometry.

Main Outcome Measures: Main outcome measures included Spearman correlation coefficients between relative afferent pupillary defect as measured by pupillometry and inter-eye differences in average RNFL thickness, VFI, and MD.

Results: Correlation coefficients between relative afferent pupillary defect and inter-eye difference in average RNFL thickness, MD, and VFI were -0.81 (P<0.001), -0.73 (P<0.001), and -0.68 (P<0.001), respectively.

Conclusions: The automated pupillometer quantifies asymmetry in afferent pupillary responses in proportion to structural and functional asymmetry as measured SD-OCT and automated static perimetry.

Download full-text PDF

Source
http://dx.doi.org/10.1097/IJG.0000000000000213DOI Listing

Publication Analysis

Top Keywords

afferent pupillary
16
measured automated
12
relative afferent
12
pupillary defect
12
rnfl thickness
12
inter-eye difference
8
difference average
8
average retinal
8
retinal nerve
8
nerve fiber
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!