Preliminary testing of GaN-based dosimeters for electron beam radiotherapy.

Radiat Prot Dosimetry

Service de Cancérologie-Radiothérapie, CHU de Grenoble, CS 10217, Université Joseph Fourier Grenoble1, Grenoble Cedex 9 F-38043, France.

Published: July 2015

The response of an implantable in vivo dosimetric system based on gallium nitride radioluminescence was investigated for electron beam radiotherapy using ELEKTA SLi and VARIAN Clinac 2100 CD Linear Accelerators. A bi-channel method has been implemented for fibre background rejection. The percentage depth dose (PDD) profiles were measured in polymethyl methacrylate for 6, 12 and 18 MeV electron beams. The PDD results were in excellent agreement with those measured with reference to ionisation chambers.