A g-factor metric for k-t SENSE and k-t PCA based parallel imaging.

Magn Reson Med

Institute for Biomedical Engineering, University and ETH Zurich, Zurich, Switzerland.

Published: February 2016

Purpose: To propose and validate a g-factor formalism for k-t SENSE, k-t PCA and related k-t methods for assessing SNR and temporal fidelity.

Methods: An analytical gxf -factor formulation in the spatiotemporal frequency domain is derived, enabling assessment of noise and depiction fidelity in both the spatial and frequency domain. Using pseudoreplica analysis of cardiac cine data the gxf -factor description is validated and example data are used to analyze the performance of k-t methods for various parameter settings.

Results: Analytical gxf -factor maps were found to agree well with pseudoreplica analysis for 3x, 5x, and 7x k-t SENSE and k-t PCA. While k-t SENSE resulted in lower average gxf values (gx (avg) ) in static regions when compared with k-t PCA, k-t PCA yielded lower gx (avg) values in dynamic regions. Temporal transfer was better preserved with k-t PCA for increasing undersampling factors.

Conclusion: The proposed gxf -factor and temporal transfer formalism allows assessing noise performance and temporal depiction fidelity of k-t methods including k-t SENSE and k-t PCA. The framework enables quantitative comparison of different k-t methods relative to frame-by-frame parallel imaging reconstruction.

Download full-text PDF

Source
http://dx.doi.org/10.1002/mrm.25606DOI Listing

Publication Analysis

Top Keywords

k-t pca
28
k-t sense
20
k-t
16
sense k-t
16
k-t methods
16
gxf -factor
16
pca k-t
12
parallel imaging
8
analytical gxf
8
frequency domain
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!