Background: Incorrect lens power remains one of the most common reasons for intraocular lens explantation/exchange. We evaluated stability and effect on capsular bag opacification of a new foldable modular lens system, with a base component and an optic component in rabbit eyes.
Methods: Bilateral phacoemulsification was performed on six rabbits; one eye received the test lens (Harmoni Modular Lens, ClarVista Medical), whereas the other received a single-piece hydrophobic acrylic control lens (SA60AT, Alcon). Slit-lamp examinations were performed at postoperative weeks 1, 2, 3, 4 and 6. All rabbits were sacrificed at week 6. After enucleation, the eyes were examined grossly from the Miyake-Apple view. Following, all globes were sectioned and processed for histopathological examination.
Results: The modular lens was stable in all six rabbits. Uveal biocompatibility was similar to or better in the test eyes when compared with control eyes up to 6 weeks postoperatively. Overall, capsular bag opacification was found to be significantly lower in the test eyes when compared with control eyes. At the 6-week gross examination, central posterior capsule opacification was scored as 0.58 ± 0.73 in the test group, and as 3.00 ± 1.26 in the control group (paired two samples for means t-test; two-tail P = 0.005).
Conclusions: The Harmoni Modular system remains stable within the capsular bag after implantation. Due to the design of the base component featuring long loops, along with peripheral anterior and posterior square edges, the lens appeared to prevent overall capsular bag opacification in the rabbit model.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1111/ceo.12526 | DOI Listing |
J Fr Ophtalmol
January 2025
Service d'ophtalmologie, centre hospitalier de Boulogne-sur-Mer, rue Jacques-Monod, 59000 Boulogne-sur-Mer, France. Electronic address:
Biomedicines
January 2025
Department of Ophthalmology, UHC Sestre Milosrdnice, 10000 Zagreb, Croatia.
We describe a novel experimental device, the VaMa (Vatavuk and Marić) artificial intraocular lens (IOL) capsule, and a method that enables all IOL types to be implanted in the bag. We present the application of the device and the procedure in patients with aphakia and native capsule damage and without efficient capsular support. The VaMa device and the method facilitate IOL exchange due to refractive errors and, in the case of their invention, the implementation of superior IOLs in the future.
View Article and Find Full Text PDFAm J Ophthalmol
January 2025
Raghudeep Eye Hospital, Laden Cataract and IOL research centre, Ahmedabad, India 380052.
Purpose: To evaluate demographic profile, risk factors and outcomes following intraocular lens (IOL) exchange surgery for late decentration of IOLs in the dead bag syndrome DESIGN: Prospective, interventional case series PARTICIPANTS: 46 eyes (43 patients) with late decentration/dislocation of IOLs and spontaneous posterior capsule rupture in a clear and relatively intact capsular bag. None of the eyes had significant zonular weakness.
Methods: Study conducted at Raghudeep Eye Hospital, India.
GMS Ophthalmol Cases
December 2024
Department of Ophthalmology, Disha Eye Hospital, Siliguri, India.
Background: Pseudophakic cystoid macular edema (CME) following primary anterior-chamber intraocular lens (ACIOL) implantations is commonly seen. Intravitreal triamcinolone acetonide (IVTA) injections have shown significant improvement in visual acuity and retinal thickness in refractory pseudophakic CME. Pseudohypopyon following IVTA injection is a known entity.
View Article and Find Full Text PDFSci Rep
December 2024
Visual Optics Lab Antwerp (VOLANTIS), Faculty of Medicine and Health Sciences, University of Antwerp, Antwerp, Belgium.
In cataract surgery, post-surgical stability of the intraocular lens plays a major role. This study aims to explore how the size and decentration of the capsulorhexis affect intraocular lens decentration and tilt by using numerical methods. Finite element models included zonules, ciliary body, capsular bag, and an IOL with two open-loop haptics were built.
View Article and Find Full Text PDFEnter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!