Ambient radiation levels in positron emission tomography/computed tomography (PET/CT) imaging center.

Radiol Bras

PhD, Titular Researcher at Centro de Desenvolvimento da Tecnologia Nuclear - Comissão Nacional de Energia Nuclear (CDTN/CNEN), Belo Horizonte, MG, Brazil.

Published: March 2015

Objective: To evaluate the level of ambient radiation in a PET/CT center.

Materials And Methods: Previously selected and calibrated TLD-100H thermoluminescent dosimeters were utilized to measure room radiation levels. During 32 days, the detectors were placed in several strategically selected points inside the PET/CT center and in adjacent buildings. After the exposure period the dosimeters were collected and processed to determine the radiation level.

Results: In none of the points selected for measurements the values exceeded the radiation dose threshold for controlled area (5 mSv/year) or free area (0.5 mSv/year) as recommended by the Brazilian regulations.

Conclusion: In the present study the authors demonstrated that the whole shielding system is appropriate and, consequently, the workers are exposed to doses below the threshold established by Brazilian standards, provided the radiation protection standards are followed.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC4366025PMC
http://dx.doi.org/10.1590/0100-3984.2013.1821DOI Listing

Publication Analysis

Top Keywords

ambient radiation
8
radiation levels
8
area msv/year
8
radiation
5
levels positron
4
positron emission
4
emission tomography/computed
4
tomography/computed tomography
4
tomography pet/ct
4
pet/ct imaging
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!