Background: To report a case series of failures of bioabsorbable interference screws with possible identification of a novel failure mechanism.

Methods: A retrospective review of ACL reconstructions by the senior author utilizing BioComposite™ Interference Screws (Arthrex, Inc., Naples, FL) was performed. Complications related to screw placement, including fracture, breakage or bending were examined. Our rate and methods of failure were compared to those quoted in the current literature.

Results: Eighty-seven patients of average age 23.8 years met inclusion criteria. There were eight screw failures in six patients, with femoral failure in seven and tibial failure in one. The femoral screw fractured halfway between the tip and head in five, while the head of the screw broke in one and the screw bent in another. In the case of tibial interference screw fracture, failure occurred halfway between the tip and head. The insertion device that was used was replaced after recognition of material deformation and considered a potential contributor to the breakage risk as no further screw failures have occurred since.

Conclusions: We demonstrate a unique failure mechanism of bio-absorbable interference screws. In each case, the reconstruction was salvaged. Regular inspection of materials and implants can ensure optimal outcomes and decrease complications intra-operatively.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.knee.2015.02.015DOI Listing

Publication Analysis

Top Keywords

interference screws
12
bioabsorbable interference
8
screw
8
interference screw
8
case series
8
screw failures
8
halfway head
8
failure
7
screw failure
4
failure anterior
4

Similar Publications

Background: Surgical techniques for biceps tenodesis vary in approach, fixation strategy, and anatomic location without clear superior technique for this common procedure.

Hypothesis/purpose: The purpose of this study was to prospectively evaluate a randomized cohort of patients undergoing arthroscopic suprapectoral (ASBT) with interference screw fixation using an inlay technique versus mini-open subpectoral (MOBT) with a unicortical button implant using an onlay technique with regards to 1) clinical outcome measures and 2) structural healing as evaluated by ultrasound.

Methods: From May 2017 to April 2021, patients undergoing biceps tenodesis were preoperatively randomized to either ASBT or MOBT.

View Article and Find Full Text PDF

Anatomic Distal Biceps Tendon Repair With All-Suture Cortical Buttons.

Arthrosc Tech

December 2024

Department of Orthopaedic Surgery, University of California Irvine, Orange, California, U.S.A.

Acute, traumatic distal biceps tendon ruptures are a common injury in the middle-aged athletic male population, with direct anatomic surgical repair being the most effective technique to restore maximal strength. Multiple techniques for distal biceps tendon repair have been described, including single- or dual-incision approaches and tendon fixation with cortical buttons, interference screws, suture anchors, and transosseous sutures. In this Technical Note, we demonstrate an anatomic distal biceps tendon repair technique with a single-incision approach using 2 all-suture cortical buttons.

View Article and Find Full Text PDF

Background: An all-inside endoscopic flexor hallucis longus (FHL) tendon transfer is indicated for the treatment of chronic, full-thickness Achilles tendon defects. The aim of this procedure is to restore function of the gastrocnemius-soleus complex while avoiding the wound complications associated with open procedures.

Description: This procedure can be performed through 2 endoscopic portals, a posteromedial portal (the working portal) and a posterolateral portal (the visualization portal).

View Article and Find Full Text PDF

Background: A lateral extra-articular tenodesis (LET) is increasingly being utilized to augment an anterior cruciate ligament reconstruction because it has been shown to reduce the risk of postreconstruction graft failure or recurrent rotatory instability. Various femoral fixation techniques are available, including the use of an interference screw, staple, or suture anchor.

Purpose: To determine and compare the biomechanical properties of an LET graft when using an interference screw, staple, or suture anchor for the femoral fixation for a modified Lemaire LET.

View Article and Find Full Text PDF

Purpose: To biomechanically evaluate a flat posterior cruciate ligament (PCL) reconstruction utilizing rectangular femoral bone tunnels.

Methods: Eight fresh-frozen human knee specimens were tested in a six-degrees-of-freedom robotic test setup. In each testing step, a force-controlled test protocol was performed, including 89 N posterior tibial translation (PTT) in neutral, internal and external rotation, from 0 to 90° of flexion.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!