Background: Surgical staplers currently all rely on the same staple form-the "B" which necessitates a high delivery profile (12 mm). A novel "D" shape staple allows for an extremely low profile of the applicator. The acute and long-term efficacy of a D-shaped staple (Cardica, Redwood City, CA, USA) was compared to conventional B-form staples (Covidien, Norwalk, CN, USA) in an animal model for intestinal transections and anastomoses.

Methods: Jejunojejunal anastomoses (JJ) were performed via mini-laparotomy in a swine model. White & blue D- and B-shaped staples were studied in three groups (planned survival 14-84 days). Intraoperative assessment included completeness of staple line, hemostasis, and need for intervention. Postoperatively, animals were evaluated for complications. At the time of sacrifice, gross pathological and histological assessments were performed.

Results: Twenty-three animals had 40 anastomoses (23 "D" and 17 "B" staple anastomoses) with no intraoperative mortalities. One "D" staple application required a manual extension of the cut. Acute hemostasis was 100%. Group 1 (n = 5) compared white staples in JJs (D staple n = 5; B staple n = 5; 14-day survival = 100%). Group 2 (n = 12) compared white staples in JJs (D staple n = 12; B staple n = 6; 34-day survival = 92 %). One animal died on day 4 for a non-staple related cause. Group 3 (n = 6) compared blue staples in JJs (D staple n = 6; B staple n = 6; 84 day survival = 84%). One animal died on day 18 due to an obstruction at the B staple JJ caused by stricture. There were no other bleeding, leaks or strictures in any of the groups. Gross pathology and histology were unremarkable in all JJs.

Conclusions: This study showed no difference in intraoperative performance and the chronic healing response in JJs between D- and B-shaped staples. Based on these findings, the D-shaped staple elicits a normal healing response in jejunostomies and offers the possibility of clinical use of this advance in staple design.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC4648953PMC
http://dx.doi.org/10.1007/s00464-015-4125-xDOI Listing

Publication Analysis

Top Keywords

staple
16
group compared
12
staples jjs
12
jjs staple
12
staple staple
12
d-shaped staple
8
b-shaped staples
8
100% group
8
compared white
8
white staples
8

Similar Publications

Maize is a staple crop worldwide, essential for food security, livestock feed, and industrial uses. Its health directly impacts agricultural productivity and economic stability. Effective detection of maize crop health is crucial for preventing disease spread and ensuring high yields.

View Article and Find Full Text PDF

Surgical Dressing Cosmesis in the Immediate Postoperative Setting: A Crowdsourcing-based Study.

Plast Reconstr Surg Glob Open

January 2025

Emory Division of Plastic and Reconstructive Surgery, Atlanta, GA.

Background: Postoperative dressings expedite wound healing and decrease the rate of infection. Options for wound dressings vary based on cost, time to apply, method of wound healing, and availability at the hospital; however, a significant difference in postoperative complications between each type has not been found. As such, this study evaluates patient cosmetic preferences for various wound dressings as it relates to early postoperative satisfaction.

View Article and Find Full Text PDF

Introduction: Post-operative complications related to hardware, particularly symptoms caused by it, such as implant impingement, wire breakage, and skin infection are frequently reported following surgical management of patellar fractures with stainless steel wires. As a potential solution, some researchers have advocated for the use of non-absorbable sutures for fracture fixation. This study aims to investigate the clinical and radiological outcomes of patients treated with fiber taping as an alternative technique.

View Article and Find Full Text PDF

The traditional standard of care, tension band wire fixation modalities commonly used to treat non-comminuted olecranon fractures, are frequently associated with complications, with symptomatic hardware being the most common issue, often necessitating subsequent surgical procedures for hardware removal. We present a case of a young, active gentleman who sustained a simple olecranon fracture (Mayo type IIA) and underwent open reduction with the innovative all-suture non-metallic internal fixation procedure. We used a low-profile, cost-effective alternative fracture fixation technique, following the principles of pre-existing surgical techniques, while utilizing FiberWire sutures with biomechanical properties equivalent to, or possibly superior to, conventionally used metallic wires, thereby eliminating the need for additional surgical intervention.

View Article and Find Full Text PDF

Objectives: To evaluate the surgical and patient-reported outcomes of YV-plasty in patients with refractory bladder neck stenosis (BNS) following transurethral prostate surgery.

Methods: This retrospective study reviewed five patients who underwent YV-plasty for BNS between January 2021 and October 2023. The surgical procedure involved a midline lower abdominal incision to expose the bladder neck.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!