Background And Purpose: A variety of ureteroscopic irrigation systems are available, ranging from gravity-driven pressure bags to hand-operated pumps. Endoscopic valves maintain a watertight seal during ureteroscopy (URS) while facilitating passage of instruments. The clinical utility and ergonomics of such devices have not been established. We systematically compare the mechanical properties and usability of select valve devices and hand-operated irrigation systems in an in vitro setting.

Materials And Methods: In vitro testing of four different endoscopic valves: UroSeal adjustable endoscopic valve (US Urology), adjustable biopsy port seal (Gyrus ACMI), Blue Silicone Seal ACMI CS B612 (Gyrus ACMI), and REF ABP Biopsy Port Seal (ACMI Corporation) was performed. Usability was evaluated via insertion/extraction forces and insertion time for instruments, including a straight tip sensor wire, 0.035″, (Boston Scientific), a laser fiber (Flexiva 200, Boston Scientific), and an Ngage Nitinol Stone Extractor 1.7F (Cook Urological) through a flexible ureteroscope (Olympus URF P5, Olympus). Flow rate, flow time, and user fatigue were tested for two irrigation systems: The single action pumping system (SAP, Boston Scientific) and the Pathfinder Plus (PP, Utah Medical Products).

Results: The US needed the shortest time for both wire insertion and basket insertion (P=0.005, and P<0.001, respectively), while the BSS needed the greatest time for laser fiber insertion (P<0.005). The REF ABP needed the greatest force for withdrawal of the Ngage basket, the laser fiber, and the Captura stone grasper through a closed seal, while the US took the least amount of force for both laser fiber withdrawal and insertion via analysis of variance. Leak point pressure assessment demonstrated that the US was leak free at irrigation pressures up to 200 mm Hg, while the ABP, BSS, and the REF ABP devices demonstrated leaks ranging from 30 to 200 mm Hg. The average and peak flow of the SAP were significantly higher than that of the PP. Mean grip strength decreased significantly after operation of the SAP for 10 minutes, while no loss of grip strength was observed after use of the PP.

Conclusions: The US valve has the advantage of facile manipulation of wires and baskets while maintaining a watertight seal, while other devices may be more cost-effective and secure. The PP has the advantage of less operator hand fatigue and ease of use, but the SAPS may allow for greater on-demand pressures. Further studies are needed to evaluate the effect of these irrigation systems on outcomes.

Download full-text PDF

Source
http://dx.doi.org/10.1089/end.2014.0891DOI Listing

Publication Analysis

Top Keywords

endoscopic valves
12
irrigation systems
12
boston scientific
12
biopsy port
8
port seal
8
gyrus acmi
8
seal acmi
8
endoscopic
4
irrigation
4
valves irrigation
4

Similar Publications

Background: Bronchoscopic lung volume reduction (BLVR) is a minimally invasive procedure used to reduce shortness of breath and improve functionality in some patients with emphysema. While BLVR is often effective for improving dyspnea by causing target lobe atelectasis, the treatment effect can sometimes be lost. This study reviews the incidence of revision bronchoscopies in patients who lost or never achieved target lobe atelectasis following BLVR.

View Article and Find Full Text PDF

Mass-forming isolated ischemic necrosis of the cecum mimicking malignancy: Clinicopathologic features of 11 cases.

Ann Diagn Pathol

December 2024

Department of Pathology and Laboratory Medicine, Indiana University School of Medicine, Indianapolis, IN, USA. Electronic address:

Isolated ischemic necrosis of the cecum (INC) is an uncommon benign condition that occurs most often in elderly patients. The clinical presentation and radiographic findings often mimic acute appendicitis or malignancy. This study aimed to investigate the clinicopathologic features of mass-forming INC.

View Article and Find Full Text PDF

Introduction: Posterior urethral valves may lead to persistent hydronephrosis (HN) and bladder dysfunction despite successful endoscopic valve ablation (EVA).

Objectives: To evaluate the effect of overnight bladder drainage (OBD) on upper urinary tracts and bladders of boys post EVA.

Materials And Methods: Boys who had persistent HN after EVA were included.

View Article and Find Full Text PDF

: Despite surgical interventions with advances in endoscopic procedures as well as shunt technologies, the quality of life in patients with hydrocephalus can be poor. Clinical experience suggests discrepancies between objective measures of treatment success and subjective patient satisfaction. With this study, we retrospectively investigated patients' knowledge of their treatment as well as their satisfaction with received interventions.

View Article and Find Full Text PDF

Intimal sarcoma of the aortic valve and ascending aorta: a case report.

J Thorac Dis

November 2024

Department of Cardiac Surgery, Beijing Anzhen Hospital, Capital Medical University, Beijing, China.

Background: Intimal sarcomas are rare tumors that typically affect the major vessels, such as the pulmonary artery and aorta, and are associated with a particularly poor prognosis. Intimal sarcomas found in the aorta are most commonly located in the abdominal section between the celiac artery and the iliac bifurcation. The descending aorta is involved in 30% of cases, while involvement of the ascending aorta is rare.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!