Background And Objectives: Radiofrequency (RF) lesions are safe and effective in the treatment of spine pain; however, models developed to study factors affecting lesion dimensions have been performed in homogeneous media that may not accurately simulate human anatomy and electrophysiology. We present a novel ex vivo porcine model for performing RF lesion studies and report the influence of bone on projection of RF ablation lesions into soft tissue.
Methods: Radiofrequency lesions were performed in porcine rib specimens using monopolar 18-gauge, 10-mm straight active tip cannula, with a lesion temperature of 80°C for 150 seconds. Ten lesions were performed in pure porcine muscle tissue and abutting porcine rib bone with surrounding muscle. Lesions were exposed with dissection and measured with digital calipers.
Results: Maximal effective lesion radius approximately doubled against the bone compared with the pure muscle group (mean, 5.65 mm [95% CI, 5.43-5.87 mm] vs 2.68 mm [95% CI, 2.55-2.81 mm], P < .0001), although this was seen only in a vertical direction and not horizontally. In addition, the prelesion and postlesion impedance of the bone-muscle interface was consistently higher than the muscle-only interface (mean, 165.6 Ohm [95% CI, 146.6-184.6 Ohm] vs 137.8 Ohm [95% CI, 135.5-140.1 Ohm], P = 0.004; 144.3 Ohm [95% CI, 134.3-154.3 Ohm] vs 124.3 Ohm [95% CI, 119.3-129.3 Ohm], P = 0.001). Other dimensions and estimated volume were not significantly different.
Conclusions: Bone adjacent to RF lesions alters the surrounding electrophysiological environment causing RF lesions to project further perpendicularly from the needle axis, vertically to bone, than previously expected. This phenomenon should be considered in the future modeling and clinical practice of RF.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1097/AAP.0000000000000221 | DOI Listing |
Pacing Clin Electrophysiol
December 2024
Department of Cardiology II - Electrophysiology, University Hospital Münster, Münster, Germany.
Background: Noninferiority of omitting intraoperative defibrillation threshold (DFT) testing has been documented for transvenous implantable cardioverter defibrillators (ICD) whereas data for the subcutaneous-ICD (S-ICD) regarding the need for DFT testing, especially during S-ICD generator replacement, is not available.
Methods: A total of 112 consecutive patients who underwent S-ICD generator replacement and routine testing were included in this retrospective single-center study and analyzed regarding the outcome of intraoperative DFT.
Results: The majority of patients (87.
Arch Sex Behav
December 2024
School of Applied Sciences for Health, Education and Social Sciences, SRH University, Merkurstr 19, 90763, Fürth, Germany.
J Anim Sci
January 2023
Department of Animal Science, University of California, Davis, CA 95616, USA.
The objectives of this study were to investigate the in vitro immune-modulatory effects of monoglycerides and zinc glycinate with porcine alveolar macrophages (PAM) and their impact on epithelial barrier integrity using the intestinal porcine enterocyte cell line (IPEC-J2). Cell viability was assessed using a Vybrant MTT assay to determine the appropriate dose range of monoglyceride blend (C4, C8, and C10) and zinc glycinate. In experiment 1, IPEC-J2 cells (5 × 105 cells/mL) were seeded and treated with each compound (monoglycerides: 0, 25, 100, 250, 500, and 1,000 µg/mL; zinc glycinate: 0, 2, 5, 12.
View Article and Find Full Text PDFPhys Rev Lett
November 2024
Centre for Space Research, North-West University, Potchefstroom 2520, South Africa.
Owing to their rapid cooling rate and hence loss-limited propagation distance, cosmic-ray electrons and positrons (CRe) at very high energies probe local cosmic-ray accelerators and provide constraints on exotic production mechanisms such as annihilation of dark matter particles. We present a high-statistics measurement of the spectrum of CRe candidate events from 0.3 to 40 TeV with the High Energy Stereoscopic System, covering 2 orders of magnitude in energy and reaching a proton rejection power of better than 10^{4}.
View Article and Find Full Text PDFJ Thromb Haemost
December 2024
Department of Medicine Solna, Karolinska Institutet, Stockholm, Sweden; Department of Haematology, Karolinska University Hospital, Stockholm Sweden.
Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!