For the first time, a multi-analyte biosensor platform has been developed using the label-free 1-lambda-reflectometry technique. This platform is the first, which does not use imaging techniques, but is able to perform multi-analyte measurements. It is designed to be portable and cost-effective and therefore allows for point-of-need testing or on-site field-testing with possible applications in diagnostics. This work highlights the application possibilities of this platform in the field of animal testing, but is also relevant and transferable to human diagnostics. The performance of the platform has been evaluated using relevant reference systems like biomarker (C-reactive protein) and serology (anti-Salmonella antibodies) as well as a panel of real samples (animal sera). The comparison of the working range and limit of detection shows no loss of performance transferring the separate assays to the multi-analyte setup. Moreover, the new multi-analyte platform allows for discrimination between sera of animals infected with different Salmonella subtypes.