Severity: Warning
Message: file_get_contents(https://...@pubfacts.com&api_key=b8daa3ad693db53b1410957c26c9a51b4908&a=1): Failed to open stream: HTTP request failed! HTTP/1.1 429 Too Many Requests
Filename: helpers/my_audit_helper.php
Line Number: 176
Backtrace:
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 176
Function: file_get_contents
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 250
Function: simplexml_load_file_from_url
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 3122
Function: getPubMedXML
File: /var/www/html/application/controllers/Detail.php
Line: 575
Function: pubMedSearch_Global
File: /var/www/html/application/controllers/Detail.php
Line: 489
Function: pubMedGetRelatedKeyword
File: /var/www/html/index.php
Line: 316
Function: require_once
Objective: To determine whether the Trifecta bioprosthetic aortic valve produces postoperative haemodynamic results comparable with or better than those of the Magna Ease aortic valve bioprosthesis.
Methods: We retrospectively reviewed the medical records of patients who had undergone aortic valve replacement with Trifecta or Magna Ease prostheses at eight European institutions between January 2011 and May 2013, and analysed early postoperative haemodynamic performance by means of echocardiography.
Results: A total of 791 patients underwent aortic valve replacement (469 Magna Ease, 322 Trifecta). Haemodynamic variables were evaluated on discharge and during the follow-up (minimum 6 months, maximum 12 months). The mean gradient and the indexed effective orifice area (IEOA) were as follows: 10 mmHg [interquartile range (IQR): 8-13] and 1.10 cm(2)/m(2) (IQR: 0.95-1.27) for Trifecta; 16 mmHg (IQR: 11-22) and 0.96 cm(2)/m(2) (IQR: 0.77-1.13) for Magna Ease (P < 0.001). These significant differences were maintained across all valve sizes. Similar statistically significant differences were found when patients were matched and/or stratified for preoperative characteristics: body-surface area, ejection fraction, mean gradients and valve size. Severe prosthesis-patient mismatch (IEOA: <0.65 cm(2)/m(2)) was detected in 2 patients (0.6%) with Trifecta and 40 patients (8.5%) with Magna Ease (P < 0.001).
Conclusions: The haemodynamic performance of the Trifecta bioprosthesis was superior to that of the Magna Ease valve across all conventional prosthesis sizes, with almost no incidence of severe patient-prosthesis mismatch. The long-term follow-up is needed to determine whether these significant haemodynamic differences will persist, and influence clinical outcomes.