Titanium mesh as a low-profile alternative for tension-band augmentation in patella fracture fixation: A biomechanical study.

Injury

Department of Orthopaedics and Rehabilitation, The University of New Mexico Health Sciences Center, MSC10 5600, 1 University of New Mexico, Albuquerque, NM 87131, United States.

Published: May 2016

Objectives: We performed a simple biomechanical study to compare the fixation strength of titanium mesh with traditional tension-band augmentation, which is a standard treatment for transverse patella fractures. We hypothesised that titanium mesh augmentation is not inferior in fixation strength to the standard treatment.

Methods: Twenty-four synthetic patellae were tested. Twelve were fixed with stainless steel wire and parallel cannulated screws. Twelve were fixed with parallel cannulated screws, augmented with anterior titanium mesh and four screws. A custom test fixture was developed to simulate a knee flexed to 90°. A uniaxial force was applied to the simulated extensor mechanism at this angle. A non-inferiority study design was used to evaluate ultimate force required for failure of each construct as a measure of fixation strength. Stiffness of the bone/implant construct, fracture gap immediately prior to failure, and modes of failure are also reported.

Results: The mean difference in force at failure was -23.0 N (95% CI: -123.6 to 77.6N) between mesh and wire constructs, well within the pre-defined non-inferiority margin of -260 N. Mean stiffness of the mesh and wire constructs were 19.42 N/mm (95% CI: 18.57-20.27 N/mm) and 19.49 N/mm (95% CI: 18.64-20.35 N/mm), respectively. Mean gap distance for the mesh constructs immediately prior to failure was 2.11 mm (95% CI: 1.35-2.88 mm) and 3.87 mm (95% CI: 2.60-5.13 mm) for wire constructs.

Conclusions: Titanium mesh augmentation is not inferior to tension-band wire augmentation when comparing ultimate force required for failure in this simplified biomechanical model. Results also indicate that stiffness of the two constructs is similar but that the mesh maintains a smaller fracture gap prior to failure. The results of this study indicate that the use of titanium mesh plating augmentation as a low-profile alternative to tension-band wiring for fixation of transverse patella fractures warrants further investigation.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.injury.2015.02.017DOI Listing

Publication Analysis

Top Keywords

titanium mesh
24
fixation strength
12
prior failure
12
mesh
9
low-profile alternative
8
alternative tension-band
8
tension-band augmentation
8
biomechanical study
8
transverse patella
8
patella fractures
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!