Background And Objectives: ABO-incompatible red blood cell transfusions still represent an important hazard in transfusion medicine. Therefore, some countries have introduced a systematic bedside ABO agglutination test checking that the right blood is given to the right patient. However, this strategy requires an extremely time-consuming learning programme and relies on a subjective interpretation of ABO test cards agglutination. We developed a prototype of a fully automated device performing the bedside agglutination test that could be completed by reading of a barcoded wristband. This POCT checks the ABO compatibility between the patient and the blood bag.
Materials And Methods: Proof of concept and analytical validation of the prototype has been completed on 451 blood samples: 238 donor packed red blood cells, 137 consecutive unselected patients for whom a blood group determination had been ordered and on 76 patient samples selected with pathology that could possibly interfere with or impair performances of the assay.
Results: We observed 100% concordance for ABO blood groups between the POCT and the laboratory instrument.
Conclusion: These preliminary results demonstrate the feasibility of ABO determination with a simple POCT device eliminating manipulation and subjective interpretation responsible for transfusion errors. This device should be linked to the blood bank system allowing all cross-check of the results.