Purpose: To objectively define the effective centration of myopic femtosecond laser-assisted LASIK ablation pattern, evaluate the difference between achieved versus planned excimer laser ablation centration, and compare these results from two different generations of an excimer laser system.
Methods: The study retrospectively evaluated 280 eyes subjected to myopic LASIK. Digital image analysis was performed on Scheimpflug sagittal curvature maps (difference of preoperative to postoperative). Centration was assessed via proprietary software digital analysis of the coordinate displacement between the achieved ablation geometric center and the planned ablation center, which was the corneal vertex. Results from two different excimer laser generations (Eye-Q 400 [140 eyes] and EX500 [140 eyes]; Alcon/WaveLight, Fort Worth, TX) were compared.
Results: Radial displacement was on average 360 ± 220 µm (range: 0 to 1,030 µm) in the Eye-Q 400 laser group and 120 ± 110 µm (range: 0 to 580 µm) in the EX500 laser group (P < .01). The percentage of eyes with displacement of greater than 300 µm was 52% in the Eye-Q 400 laser group and 4% in the EX500 laser group.
Conclusions: Displacement of ablation pattern may depend on the laser platform used. The improvement in the efficiency of centration indicates that newer generation excimer lasers with faster eye tracking and active centration control appear to achieve a significantly more accurate centration of myopic ablation patterns. The authors propose this novel, objective technique for laser refractive surgeon evaluation may point out significant outcome measures not currently used in standard metrics of refractive laser efficiency.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.3928/1081597X-20150225-01 | DOI Listing |
Arch Dermatol Res
January 2025
Department of Physiotherapy, Institute of Dermatology, Chinese Academy of Medical Sciences & Peking Union Medical College, Nanjing, 210042, China.
Background: Previous studies have proven that 308-nm light-emitting diode(308-nm LED)and 308-nm excimer lamp(308-nm MEL) are effective in treating vitiligo, but there is a lack of comparison of their efficacy for facial lesions.
Objective: To evaluate and contrast the treatment success rates of 308-nm LED versus 308-nm excimer lamp in managing facial lesions among patients suffering from stable non-segmental vitiligo.
Methods: The enrolled 119 patients with 145 lesions were randomly assigned to receive 308-nm LED or 308-nm MEL for two months.
Purpose: To evaluate the clinical outcome of laser-assisted surgical correction of high hyperopic or mixed astigmatism using small incision intrastromal lenticule rotation (SMILERO) alone or combined with photorefractive keratectomy (PRK).
Methods: This retrospective case series enrolled 25 eyes with high astigmatism that underwent SMILERO surgery. Uncorrected distance visual acuity (UDVA), corrected distance visual acuity (CDVA), uncorrected near visual acuity (UNVA), manifest refraction, central corneal thickness, and corneal higher order aberrations were analyzed before surgery and after 3, 6, and 12 months of follow-up.
Purpose: To evaluate a new regression-derived nomogram for high myopic astigmatism in small incision lenticule extraction (SMILE) surgery.
Methods: In this prospective study, data from 180 eyes with myopic astigmatism ranging from -2.50 to -4.
Purpose: To compare the clinical outcomes, surgical workflow, and patient satisfaction following small incision lenticule extraction (SMILE) performed with the VisuMax 800 in one eye and the VisuMax 500 in the contralateral eye (both Carl Zeiss Meditec).
Methods: This was a prospective, single-site clinical study of patients undergoing SMILE for myopia and myopic astigmatism between February 2022 and August 2023. Each patient underwent bilateral treatment using the VisuMax 800 (VM800 group) in one eye and the VisuMax 500 (VM500 group) in the contralateral eye.
Purpose: To investigate long-term corneal biomechanical changes in thin corneas after small incision lenticule extraction (SMILE).
Methods: Patients with indications for SMILE were enrolled in this study between November 2017 and March 2018. Patients were matched for age, spherical diopter, cylinder, spherical equivalent (SE), and lenticule thickness (LT), and then categorized into the thin cornea group (preoperative thinnest central corneal thickness [CCT] of 500 µm or less, 32 eyes) or normal cornea group (CCT of greater than 500 µm, 32 eyes).