Objective: Undertaking of a complete audit of the service of mammography, as recommended by BI-RADS(®), in a private reference institution for breast cancer diagnosis in the city of São Paulo, SP, Brazil, and comparison of results with those recommended by the literature.
Materials And Methods: Retrospective, analytical and cross-sectional study including 8,000 patients submitted to mammography in the period between April 2010 and March 2011, whose results were subjected to an internal audit. The patients were followed-up until December 2012.
Results: The radiological classification of 7,249 screening mammograms, according to BI-RADS, was the following: category 0 (1.43%), 1 (7.82%), 2 (80.76%), 3 (8.35%), 4 (1.46%), 5 (0.15%) and 6 (0.03%). The breast cancer detection ratio was 4.8 cases per 1,000 mammograms. Ductal carcinoma in situ was found in 22.8% of cases. Positive predictive values for categories 3, 4 and 5 were 1.3%, 41.3% and 100%, respectively. In the present study, the sensitivity of the method was 97.1% and specificity, 97.4%.
Conclusion: The complete internal audit of a service of mammography is essential to evaluate the quality of such service, which reflects on an early breast cancer detection and reduction of mortality rates.