Purpose: (18)F-Labelled fluorodeoxyglucose (FDG) can detect early changes in tumour metabolism and may be a useful quantitative imaging biomarker (QIB) for prediction of disease stabilization, response and duration of progression-free survival (PFS). Standardization of imaging procedures is a prerequisite, especially in multicentre clinical trials. In this study we reviewed the quality of FDG scans and compliance with the imaging guideline (IG) in a phase III clinical trial.

Methods: Forty-four cancer patients were enroled in an imaging sub-study of a randomized international multicentre trial. FDG scan had to be performed at baseline and 10-14 days after treatment start. The image transmittal forms (ITFs) and Digital Imaging and Communications in Medicine (DICOM) [1] standard headers were analysed for compliance with the IG. Mean liver standardized uptake values (LSUVmean) were measured as recommended by positron emission tomography (PET) Response Criteria in Solid Tumors 1.0 (PERCIST) [2].

Results: Of 88 scans, 81 were received (44 patients); 36 were properly anonymized; 77/81 serum glucose values submitted, all but one within the IG. In 35/44 patients both scans were of sufficient visual quality. In 22/70 ITFs the reported UT differed by >1 min from the DICOM headers (max. difference 1 h 4 min). Based on the DICOM, UT compliance for both scans was 31.4%. LSUVmean was fairly constant for the 11 patients with UT compliance: 2.30 ± 0.33 at baseline and 2.27 ± 0.48 at follow-up (FU). Variability substantially increased for the subjects with unacceptable UT (11 patients): 2.27 ± 1.04 at baseline and 2.18 ± 0.83 at FU.

Conclusion: The high attrition number of patients due to low compliance with the IG compromised the quantitative assessment of the predictive value for early response monitoring. This emphasizes the need for better regulated procedures in imaging departments, which may be achieved by education of involved personnel or efforts towards regulations. LSUVmean could be monitored to assess quality and compliance in an FDG PET/CT study.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC4382532PMC
http://dx.doi.org/10.1007/s00259-015-3002-0DOI Listing

Publication Analysis

Top Keywords

fdg scans
8
imaging sub-study
8
phase iii
8
imaging
7
compliance
6
patients
6
fdg
5
scans
5
retrospective quality
4
quality control
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!