A novel tribological study on DLC-coated micro-dimpled orthopedics implant interface.

J Mech Behav Biomed Mater

Faculty of Mechanical Engineering, Brno University of Technology, Technická 2896/2, 616 69 Brno, Czech Republic; CEITEC-Central European Institute of Technology, Brno University of Technology, Technická 3058/10, 616 00 Brno, Czech Republic.

Published: May 2015

This study investigates a tribological performance of diamond like carbon (DLC) coated micro dimpled prosthesis heads against ceramic cups in a novel pendulum hip joint simulator. The simulator enables determining friction coefficient and viscous effects of a concave shaped specimen interface (conformal contact). Two types of DLC such as hydrogenated amorphous carbon (a-C:H) and tetrahedral amorphous carbon (Ta-C) and one set of micro dimple (diameter of 300µm, depth of 70µm, and pitch of 900µm) were fabricated on metallic prosthesis heads. The experiment results reveal a significant friction coefficient reduction to the 'dimpled a-C:H/ceramic' prosthesis compared to a 'Metal (CoCr)/ceramic' prosthesis because of their improved material and surface properties and viscous effect. The post-experiment surface analysis displays that the dimpled a-C:H yielded a minor change in the surface roughness, and generated a larger sizes of wear debris (40-200nm sized, equivalent diameter), a size which could be certainly stored in the dimple, thus likely to reducing their possible third body abrasive wear rate. Thus, dimpled a:C-H can be used as a 'metal on ceramic hip joint interface', whereas the simulator can be utilized as an advanced bio-tribometer.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.jmbbm.2014.11.028DOI Listing

Publication Analysis

Top Keywords

prosthesis heads
8
hip joint
8
friction coefficient
8
amorphous carbon
8
novel tribological
4
tribological study
4
study dlc-coated
4
dlc-coated micro-dimpled
4
micro-dimpled orthopedics
4
orthopedics implant
4

Similar Publications

Comparison of bioabsorbable screw versus metallic screw fixation for tibial tubercle fractures in adolescents: a retrospective cohort study.

BMC Musculoskelet Disord

January 2025

Department of Orthopedics, Beijing Children's Hospital, Capital Medical University, National Center for Children's Health, Beijing, 100045, China.

Background: Displaced tibial tubercle (TT) fractures in adolescents are typically treated with open reduction and internal fixation. While metallic screw (MS) fixation provides strong stability, it often results in a high incidence of postoperative screw head protrusion or irritation, leading to additional removal surgery. Bioabsorbable screw (BS) fixation presents an alternative that may avoid these issues, though its stability has not yet been extensively documented in the literature.

View Article and Find Full Text PDF

Computer-aided design and fabrication of statistically shaped nasal prostheses: A feasibility study.

J Prosthet Dent

January 2025

Head and Neck Surgeon and Head, Verwelius 3D Lab, Department of Head and Neck Oncology and Surgery, Netherlands Cancer Institute, Amsterdam, The Netherlands.

Statement Of Problem: A nasal prosthesis may compensate for a partial or complete defect of the nose associated with trauma or amputation. However, the design and production is time-consuming, expensive, and expertize-dependent. Computer-generated prosthesis models and 3D printing can optimize the process.

View Article and Find Full Text PDF

Objective: To compare the 3-year outcomes of the modified minimally invasive Ponto surgery (m-MIPS) to both the original MIPS (o-MIPS) and linear incision technique with soft tissue preservation (LIT-TP) for inserting bone-anchored hearing implants (BAHIs).

Study Design: Prospective study with three patient groups: m-MIPS, o-MIPS, and LIT-TP.

Setting: Tertiary referral center.

View Article and Find Full Text PDF

Detection of Extracochlear Electrodes Using Electrical Field Imaging (EFI).

Otol Neurotol

February 2025

Department of Otolaryngology-Head and Neck Surgery, Mayo Clinic, Rochester, Minnesota.

Objective: To analyze the use of electrical field imaging (EFI) in the detection of extracochlear electrodes in cochlear implants (CI).

Study Design: Retrospective cohort study.

Setting: Tertiary academic medical center.

View Article and Find Full Text PDF

Objective: The aim of this study is to test the feasibility of a custom 3D-printed guide for performing a minimally invasive cochleostomy for cochlear implantation.

Study Design: Prospective performance study.

Setting: Secondary care.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!