Significant pressure differences between solid-state and water-perfused systems in lower esophageal sphincter measurement.

Surg Endosc

Center of Operative Medicine, Department of Visceral, Transplant and Thoracic Surgery, Medical University Innsbruck, Anichstrasse 35, 6020, Innsbruck, Austria.

Published: December 2015

Objective: High-resolution manometry of the esophagus has gained worldwide acceptance, using different solid-state catheters. Thus, normal values for lower esophageal sphincter (LES) resting pressure in suspected gastroesophageal reflux disease patients have been established using water-perfused manometry. These standard values are commonly applied using also solid-state techniques, although they have never been compared before. The aim of the study was to compare LES measurements obtained with water-perfused manometry with a solid-state technique.

Methods: Thirty healthy subjects were studied twice on the same day: Technique 1: Station pull through using a water-perfused catheter with ports arranged at 0°, 90°, 180° and 270° which were averaged to give a mean LES pressure. Technique 2: Solid-state circumferential probe with a single station pull through. Data were collected using the same computer system and program. The LES pressures were randomly and blindly analyzed.

Results: Twenty-seven subjects out of 30 were analyzed. Using the solid-state system, the mean LES pressure was higher (15.0 vs. 23.3 mmHg, p = 0.003) and 19 of 27 (70%) individual measurements were higher. Two subjects had a hypertensive LES by solid state (58.6 resp. 47.5 mmHg), while their pressures were normal with water-perfused manometry (21.0 resp. 23.4 mmHg). The distal esophageal pressures (mean of pressure at 3 and 8 cm above LES) were the same with the two techniques.

Conclusion: In normal control subjects, LES measurement using circumferential solid-state transducers yields higher pressures than standard water-perfused manometry. Which system yields the "true" resting pressure of the physiologic LES remains to be determined.

Download full-text PDF

Source
http://dx.doi.org/10.1007/s00464-015-4109-xDOI Listing

Publication Analysis

Top Keywords

water-perfused manometry
16
les
9
lower esophageal
8
esophageal sphincter
8
resting pressure
8
station pull
8
les pressure
8
solid-state
7
pressure
6
water-perfused
6

Similar Publications

Background: The current definition for hypercontractile esophagus was arbitrarily set at the uppermost range in volunteers for a specific equipment.

Objective: This study aims to critically analyze the concept of hypercontractile waves to redefine hypercontractile esophagus parameters.

Methods: We reviewed 500 unselected and consecutive HRM tests (5000 waves) performed in a water -perfused system.

View Article and Find Full Text PDF

INFLUENCE OF ESOPHAGEAL MOTILITY IMPAIRMENT ON UPPER AND LOWER ESOPHAGEAL SPHINCTER PRESSURE IN CHAGAS DISEASE.

Arq Gastroenterol

June 2024

Faculdade de Medicina de Ribeirão Preto, Universidade de São Paulo, Ribeirão Preto SP, Brasil.

Background: Chagas disease causes digestive anatomic and functional changes, including the loss of the myenteric plexus and abnormal esophageal radiologic and manometric findings.

Objective: To evaluate the association of abnormal esophageal radiologic findings, cardiac changes, distal esophageal contractions, and complaints of dysphagia and constipation in upper (UES) and lower (LES) esophageal sphincter basal pressure in Chagas disease patients.

Methods: The study evaluated 99 patients with Chagas disease and 40 asymptomatic normal volunteers.

View Article and Find Full Text PDF

Background/aim: London Protocol (LP) and Classification allied to high-resolution manometry (HRM) technological evolution has updated and enhanced the diagnostic armamentarium in anorectal disorders. This study aims to evaluate LP reproducibility under water-perfused HRM, provide normal data and new parameters based on 3D and healthy comparison studies under perfusional HRM.

Methods: Fifty healthy (25 F) underwent water-perfused 36 channel HRM based on LP at resting, squeeze, cough, push, and rectal sensory.

View Article and Find Full Text PDF

Background/purpose: We investigated the diagnostic performance of the anal sphincter relaxation integral (ASRI) for infants with Hirschsprung's disease (HD).

Methods: We performed water-perfused high-resolution anorectal manometry (HRAM) in 18 infants (9 with HD), and solid-state HRAM in another 18 infants (4 with HD). We calculated the ASRI during the rectoanal inhibitory reflex (RAIR) maneuver at pressure cutoffs of <10 mmHg (ASRI 10) and <15 mmHg (ASRI 15).

View Article and Find Full Text PDF

Background: Anorectal manometry is a valuable tool for objective assessment of motility motor and sensory function of the anorectum. The aim of this study was to investigate role of water-perfused (WP) three dimensional high-resolution anorectal manometry (3D-HRARM) compared to WP conventional manometry (CM) in the management of chronic idiopathic constipation (CIC) and faecal incontinence (FI) in children.

Methods: This was a retrospective review of 122 consecutive children, who had WP 8-channel CM or 24 channel 3D-HRARM and endosonography under ketamine anaesthesia from September 2012 to February 2019.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!