Background: Electromagnetic fields and ionising radiation during radiotherapy can influence the functioning of ICDs. Guidelines for radiotherapy treatment were published in 1994, but only based on experience with pacemakers. Data on the influence of radiotherapy on ICDs is limited.
Objectives: We determined the risk to ICDs of interference detection induced by radiotherapy.
Methods: In our study we irradiated 11 ICDs. The irradiation was performed with a 6 megavolt photon beam. In each individual device test, a total of 20 Gray was delivered in a fractionated fashion. During each irradiation the output stimulation rate was monitored and electrogram storage was activated. In case of interference the test was repeated with the ICD outside and the lead(s) inside and outside the irradiation field.
Results: With the ICD inside the irradiation field, interference detection was observed in all ICDs. This caused pacing inhibition or rapid ventricular pacing. Ventricular tachycardia (VT) or ventricular fibrillation (VF) detection occurred, which would have caused tachycardia-terminating therapy. If the ICD was placed outside the irradiation field, no interference was observed.
Conclusion: Interference by ionising radiation on the ICDs is demonstrated both on bradycardia and tachycardia therapy. This can have consequences for patients. Recommendations for radiotherapy are presented in this article.
Download full-text PDF |
Source |
---|---|
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC2557320 | PMC |
Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!