Goals: To compare the quality of low-dose CT images with sinogram affirmed iterative reconstruction (SAFIRE), and full-dose CT with filtered back projection reconstructions (FBP).
Materials And Methods: Fifty pulmonary CT performed by a dual-source technique (120kVp; 110mAs) with (a) the same energy in both tubes, and (b) the distribution of reference mAs with 40% in tube A (44mAs) and 60% in tube B (66mAs). Each acquisition allowed reconstruction of: (a) full-dose images (with both tubes) with FBP reconstructions (group 1); and (b) low-dose images (from tube A) reconstructed with SAFIRE (group 2).
Results: Group 2 images presented: (a) a significant objective reduction in noise measured in the trachea on mediastinal (16.04±5.66 vs 17.66±5.84) (P=0.0284) and pulmonary (29.77±6.79 vs 37.96±9.03) (P<0.0001) images; (b) a similar subjective perception of noise and overall image quality (P=1), which was considered to be excellent in 66% (33/50) of the cases, with no influence on the detection of elementary pulmonary lesions of infiltration (98.4%; 95% CI=[96.9%-99.9%]).
Conclusion: Despite a 60% reduction in radiation dose, the image quality with iterative reconstruction is objectively better and subjectively similar to full-dose FBP images.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1016/j.diii.2014.08.006 | DOI Listing |
Med Phys
January 2025
Faculty of Physics, Astronomy and Applied Computer Science, Jagiellonian University, Kraków, Poland.
Background: Total-body (TB) Positron Emission Tomography (PET) is one of the most promising medical diagnostics modalities, opening new perspectives for personalized medicine, low-dose imaging, multi-organ dynamic imaging or kinetic modeling. The high sensitivity provided by total-body technology can be advantageous for novel tomography methods like positronium imaging, demanding the registration of triple coincidences. Currently, state-of-the-art PET scanners use inorganic scintillators.
View Article and Find Full Text PDFMater Today Bio
February 2025
Kunming Institute of Zoology, Chinese Academy of Sciences, PR China.
The early treatment of Osteonecrosis of Femoral Head (ONFH) remains a clinical challenge. Conventional Bone Marrow Mesenchymal Stem Cell (BMSC) injection methods often result in unsatisfactory outcomes due to mechanical cell damage, low cell survival and retention rates, inadequate cell matrix accumulation, and poor intercellular interaction. In this study, we employed a novel cell carrier material termed "3D Microscaffold" to deliver BMSCs, addressing these issues and enhancing the therapeutic effects of cell therapy for ONFH.
View Article and Find Full Text PDFBMC Cardiovasc Disord
January 2025
Department of Computed Tomography and Magnetic Resonance, Fourth Hospital of Hebei Medical University, Shijiazhuang, China.
Objectives: This study aimed to evaluate the feasibility and accuracy of non-electrocardiogram (ECG)-triggered chest low-dose computed tomography (LDCT) with a kV-independent reconstruction algorithm in assessing coronary artery calcification (CAC) degree and cardiovascular disease risk in patients receiving maintenance hemodialysis (MHD).
Methods: In total, 181 patients receiving MHD who needed chest CT and coronary artery calcium score (CACS) scannings sequentially underwent non-ECG-triggered, automated tube voltage selection, high-pitch chest LDCT with a kV-independent reconstruction algorithm and ECG-triggered standard CACS scannings. Then, the image quality, radiation doses, Agatston scores (ASs), and cardiac risk classifications of the two scans were compared.
Jpn J Radiol
January 2025
Department of Radiology, Ehime University Graduate School of Medicine, Shitsukawa, Toon, Ehime, 791-0295, Japan.
Purpose: To evaluate the effects of four-dimensional noise reduction filtering using a similarity algorithm (4D-SF) on the image quality and tumor visibility of low-dose dynamic computed tomography (CT) in evaluating breast cancer.
Materials And Methods: Thirty-four patients with 38 lesions who underwent low-dose dynamic breast CT and were pathologically diagnosed with breast cancer were enrolled. Dynamic CT images were reconstructed using iterative reconstruction alone or in combination with 4D-SF.
J Imaging Inform Med
January 2025
School of Electronic and Information Engineering, Beijing Jiaotong University, Beijing, China.
While radiation hazards induced by cone-beam computed tomography (CBCT) in image-guided radiotherapy (IGRT) can be reduced by sparse-view sampling, the image quality is inevitably degraded. We propose a deep learning-based multi-view projection synthesis (DLMPS) approach to improve the quality of sparse-view low-dose CBCT images. In the proposed DLMPS approach, linear interpolation was first applied to sparse-view projections and the projections were rearranged into sinograms; these sinograms were processed with a sinogram restoration model and then rearranged back into projections.
View Article and Find Full Text PDFEnter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!