CT pulmonary angiogram with 60% dose reduction: Influence of iterative reconstructions on image quality.

Diagn Interv Imaging

Service d'Imagerie Thoracique, Hôpital Calmette (EA 2694), CHRU et Université de Lille 2 Nord de France, boulevard Jules-Leclercq, 59000 Lille, France. Electronic address:

Published: May 2015

Goals: To compare the quality of low-dose CT images with sinogram affirmed iterative reconstruction (SAFIRE), and full-dose CT with filtered back projection reconstructions (FBP).

Materials And Methods: Fifty pulmonary CT performed by a dual-source technique (120kVp; 110mAs) with (a) the same energy in both tubes, and (b) the distribution of reference mAs with 40% in tube A (44mAs) and 60% in tube B (66mAs). Each acquisition allowed reconstruction of: (a) full-dose images (with both tubes) with FBP reconstructions (group 1); and (b) low-dose images (from tube A) reconstructed with SAFIRE (group 2).

Results: Group 2 images presented: (a) a significant objective reduction in noise measured in the trachea on mediastinal (16.04±5.66 vs 17.66±5.84) (P=0.0284) and pulmonary (29.77±6.79 vs 37.96±9.03) (P<0.0001) images; (b) a similar subjective perception of noise and overall image quality (P=1), which was considered to be excellent in 66% (33/50) of the cases, with no influence on the detection of elementary pulmonary lesions of infiltration (98.4%; 95% CI=[96.9%-99.9%]).

Conclusion: Despite a 60% reduction in radiation dose, the image quality with iterative reconstruction is objectively better and subjectively similar to full-dose FBP images.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.diii.2014.08.006DOI Listing

Publication Analysis

Top Keywords

low-dose images
8
pulmonary angiogram
4
angiogram 60%
4
60% dose
4
dose reduction
4
reduction influence
4
influence iterative
4
iterative reconstructions
4
reconstructions image
4
image quality
4

Similar Publications

Background: Total-body (TB) Positron Emission Tomography (PET) is one of the most promising medical diagnostics modalities, opening new perspectives for personalized medicine, low-dose imaging, multi-organ dynamic imaging or kinetic modeling. The high sensitivity provided by total-body technology can be advantageous for novel tomography methods like positronium imaging, demanding the registration of triple coincidences. Currently, state-of-the-art PET scanners use inorganic scintillators.

View Article and Find Full Text PDF

The early treatment of Osteonecrosis of Femoral Head (ONFH) remains a clinical challenge. Conventional Bone Marrow Mesenchymal Stem Cell (BMSC) injection methods often result in unsatisfactory outcomes due to mechanical cell damage, low cell survival and retention rates, inadequate cell matrix accumulation, and poor intercellular interaction. In this study, we employed a novel cell carrier material termed "3D Microscaffold" to deliver BMSCs, addressing these issues and enhancing the therapeutic effects of cell therapy for ONFH.

View Article and Find Full Text PDF

Objectives: This study aimed to evaluate the feasibility and accuracy of non-electrocardiogram (ECG)-triggered chest low-dose computed tomography (LDCT) with a kV-independent reconstruction algorithm in assessing coronary artery calcification (CAC) degree and cardiovascular disease risk in patients receiving maintenance hemodialysis (MHD).

Methods: In total, 181 patients receiving MHD who needed chest CT and coronary artery calcium score (CACS) scannings sequentially underwent non-ECG-triggered, automated tube voltage selection, high-pitch chest LDCT with a kV-independent reconstruction algorithm and ECG-triggered standard CACS scannings. Then, the image quality, radiation doses, Agatston scores (ASs), and cardiac risk classifications of the two scans were compared.

View Article and Find Full Text PDF

Purpose: To evaluate the effects of four-dimensional noise reduction filtering using a similarity algorithm (4D-SF) on the image quality and tumor visibility of low-dose dynamic computed tomography (CT) in evaluating breast cancer.

Materials And Methods: Thirty-four patients with 38 lesions who underwent low-dose dynamic breast CT and were pathologically diagnosed with breast cancer were enrolled. Dynamic CT images were reconstructed using iterative reconstruction alone or in combination with 4D-SF.

View Article and Find Full Text PDF

While radiation hazards induced by cone-beam computed tomography (CBCT) in image-guided radiotherapy (IGRT) can be reduced by sparse-view sampling, the image quality is inevitably degraded. We propose a deep learning-based multi-view projection synthesis (DLMPS) approach to improve the quality of sparse-view low-dose CBCT images. In the proposed DLMPS approach, linear interpolation was first applied to sparse-view projections and the projections were rearranged into sinograms; these sinograms were processed with a sinogram restoration model and then rearranged back into projections.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!