Severity: Warning
Message: file_get_contents(https://...@gmail.com&api_key=61f08fa0b96a73de8c900d749fcb997acc09&a=1): Failed to open stream: HTTP request failed! HTTP/1.1 429 Too Many Requests
Filename: helpers/my_audit_helper.php
Line Number: 176
Backtrace:
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 176
Function: file_get_contents
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 250
Function: simplexml_load_file_from_url
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 1034
Function: getPubMedXML
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 3152
Function: GetPubMedArticleOutput_2016
File: /var/www/html/application/controllers/Detail.php
Line: 575
Function: pubMedSearch_Global
File: /var/www/html/application/controllers/Detail.php
Line: 489
Function: pubMedGetRelatedKeyword
File: /var/www/html/index.php
Line: 316
Function: require_once
Background: Although interest in sphincter-sparing treatments for anal fistulas is increasing, few large prospective studies of these approaches have been conducted.
Objective: The study assessed outcomes after implantation of a synthetic bioabsorbable anal fistula plug.
Design: A prospective, multicenter investigation was performed.
Setting: The study was conducted at 11 colon and rectal centers.
Patients: Ninety-three patients (71 men; mean age, 47 years) with complex cryptoglandular transsphincteric anal fistulas were enrolled. Exclusion criteria included Crohn's disease, an active infection, a multitract fistula, and an immunocompromised status.
Intervention: Draining setons were used at the surgeon's discretion. Patients had follow-up evaluations at 1, 3, 6, and 12 months postoperatively.
Main Outcome Measures: The primary end point was healing of the fistula, defined as drainage cessation plus closure of the external opening, at 6 and 12 months. Secondary end points were fecal continence, duration of drainage from the fistula, pain, and adverse events during follow-up.
Results: Thirteen patients were lost to follow-up and 21 were withdrawn, primarily to undergo an alternative treatment. The fistula healing rates at 6 and 12 months were 41% (95% CI, 30%-52%; total n = 74) and 49% (95% CI, 38%-61%; total n = 73). Half the patients in whom a previous treatment failed had healing. By 6 months, the mean Wexner score had improved significantly (p = 0.0003). By 12 months, 93% of patients had no or minimal pain. Adverse events included 11 infections/abscesses, 2 new fistulas, and 8 total and 5 partial plug extrusions. The fistula healed in 3 patients with a partial extrusion.
Limitations: The study was nonrandomized and had relatively high rates of loss to follow-up.
Conclusion: Implantation of a synthetic bioabsorbable fistula plug is a reasonably efficacious treatment for complex transsphincteric anal fistulas, especially given the simplicity and low morbidity of the procedure.