AI Article Synopsis

  • The study tested a new double-lumen endotracheal tube (DLT) called VivaSight, which has a built-in camera for better visibility during intubation, against a conventional DLT in thoracic surgery patients.
  • The results showed that the VivaSight DLT allowed for faster intubation times (63 seconds on average) compared to the conventional DLT (97 seconds), and it was successfully inserted in all attempts.
  • Although it had a slightly higher incidence of airway injuries, the VivaSight DLT proved to be easier to position and adjust, confirming its potential advantages in surgical settings.

Article Abstract

Introduction: Double-lumen endotracheal tubes (DLTs), which are commonly used for single-lung ventilation during surgery, are difficult to insert. In addition, they often move during surgical lung manipulation which can cause life-threatening complications. Flexible bronchoscopy is used routinely to establish and confirm proper DLT placement. The newly designed VivaSight DLT has an integrated camera, allowing continuous visualization of its position in the trachea. We hypothesized that the time to intubation using the VivaSight DLT would be faster than with a conventional DLT.

Methods: We enrolled 40 adults scheduled for thoracic surgery. Patients were randomized to conventional DLT (n = 20) or VivaSight DLT (n = 20). Time to intubation was our primary outcome. Secondary outcomes were insertion success without flexible bronchoscopy, frequency of tube displacement, ease of insertion, quality of lung collapse, postoperative complaints, and airway injuries.

Results: Time [mean (SD)] to successful intubation was significantly faster with the VivaSight DLT [63 (58) sec] compared with the conventional DLT [97 (84) sec; P = 0.03]. The VivaSight DLTs were correctly inserted during all attempts. When malpositioning of the VivaSight DLT occurred, it was easily remedied, even in the lateral position. The devices were comparable with respect to postoperative coughing, hoarseness, and sore throat. Airway injuries tended to be more common with the VivaSight DLT, although this study was underpowered for airway injuries.

Conclusion: The VivaSight DLT camera allowed faster insertion and facilitated initial positioning. It also confirmed proper tube positioning intraoperatively and facilitated repositioning when necessary. This trial was registered at clinicaltrials.gov: NCT01807676.

Download full-text PDF

Source
http://dx.doi.org/10.1007/s12630-015-0329-8DOI Listing

Publication Analysis

Top Keywords

vivasight dlt
28
dlt
10
intubation vivasight
8
flexible bronchoscopy
8
vivasight
8
time intubation
8
conventional dlt
8
dlt n = 20
8
intubation
4
conventional
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!