Purpose: To determine whether an unmodified commercial wavefront aberrometer (irx3) can be used to estimate forward light scattering and how this assessment matches estimations obtained from the C-Quant straylight meter.

Setting: University of Manchester, Manchester, United Kingdom.

Design: Prospective comparative study.

Methods: Measurements obtained with a straylight meter and with Shack-Hartmann spot patterns using a previously reported metric were compared. The method was first validated in a model eye by spraying an aerosol over 4 contact lenses to generate various levels of scattering. Measurements with both methods were subsequently obtained in healthy eyes.

Results: The study comprised 33 healthy participants (mean age 38.9 years ± 13.1 [SD]). A good correlation was observed between the density of droplets over the contact lenses and the objective scatter value extracted from the hartmanngrams (r = 0.972, P < .05). However, in eyes of subjects, no significant relationship was found between measurements from the straylight meter and the metric derived from the Shack-Hartmann method (r = 0.133, P = .460).

Conclusions: The hartmanngrams provided a valid objective measurement of the light scatter in a model eye; the measurements in human eyes were not significantly correlated with those of the light scatter meter. The straylight meter assesses large-angle scattering, while the Shack-Hartmann method collates information from a narrow angle around the center of the point-spread function; this could be the reason for the difference in measurements.

Financial Disclosure: No author has a financial or proprietary interest in any material or method mentioned.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.jcrs.2014.04.042DOI Listing

Publication Analysis

Top Keywords

straylight meter
16
light scatter
12
forward light
8
shack-hartmann spot
8
spot patterns
8
measurements straylight
8
model eye
8
contact lenses
8
shack-hartmann method
8
straylight
5

Similar Publications

Subjective Straylight Index: A Visual Test for Retinal Contrast Assessment as a Function of Veiling Glare.

J Imaging

April 2024

Departamento de Física Aplicada, Facultad de Ciencias, Universidad de Zaragoza, 50009 Zaragoza, Spain.

Spatial aspects of visual performance are usually evaluated through visual acuity charts and contrast sensitivity (CS) tests. CS tests are generated by vanishing the contrast level of the visual charts. However, the quality of retinal images can be affected by both ocular aberrations and scattering effects and none of those factors are incorporated as parameters in visual tests in clinical practice.

View Article and Find Full Text PDF

The screen image synthesis (SIS) meter was originally proposed as a high-speed measurement tool, which fused the measured data from multiple sample-rotational angles to produce a whole-field measurement result. However, it suffered from stray light noise and lacked the capability of spectrum measurement. In this study, we propose an SIS system embedded with a snapshot hyperspectral technology, which was based on a dispersion image of the sparse sampling screen (SSS).

View Article and Find Full Text PDF

Changes in straylight after cataract surgery.

J Cataract Refract Surg

March 2024

From the Department of Ophthalmology, Amphia Hospital, Breda, the Netherlands (Reus); Netherlands Institute for Neuroscience, Royal Netherlands Academy of Arts and Sciences, Amsterdam, the Netherlands (van den Berg).

Purpose: To investigate straylight in the immediate postoperative period after cataract surgery.

Setting: Amphia Hospital, Breda, the Netherlands.

Design: Prospective, comparative, single-arm, single-center, single-surgeon study.

View Article and Find Full Text PDF

Forward-scattered and backward-scattered light in moderate keratoconus.

Ophthalmic Physiol Opt

July 2023

Optometry and Vision Department, Faculty of Optics and Optometry, Universidad Complutense de Madrid, Madrid, Spain.

Introduction: To evaluate the backscattered light, objective scatter index (OSI) and retinal straylight in patients with moderate keratoconus and healthy control subjects.

Methods: A prospective observational study was developed with 33 patients in the moderate-keratoconus group (KC) and 34 in the non-keratoconus group (NKC). Corneal densitometry was obtained using Scheimpflug corneal tomography and measurements were expressed in grayscale units (GSU) over four zones within a 12.

View Article and Find Full Text PDF

Purpose: To evaluate the relationship between intraocular straylight perception and: (i) contrast sensitivity (CS), (ii) halo size, and (iii) hazard recognition distance, in the presence and absence of glare.

Subjects And Methods: Participants were 15 (5 female) ophthalmologically healthy adults, aged 54.6-80.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!