Automated detection of patient movement during a CBCT scan based on the projection data.

Oral Surg Oral Med Oral Pathol Oral Radiol

Computing for New Media, RheinMain University of Applied Sciences, Wiesbaden, Germany.

Published: April 2015

Objectives: To develop an automated procedure to detect patient motion on the projection images acquired during a cone beam computed tomography (CBCT) scan and to evaluate the method's feasibility on small real-world CBCT images in relation to visual assessment.

Methods: Based on optical flow theory, software was developed using the sequence of the projection images of a CBCT machine for automated detection of patient motion. Averaged acceleration vectors were used as measurement data and compared with visual assessment of the projection images displayed as video. Seventy-nine CBCT data sets (small field-of-view: 40 mm) from our patient database were selected in a sequential fashion and evaluated with the software.

Results: 10 out of 79 (13%) were allocated to a patient movement. A threshold of 0.4 pixel/frame transition was empirically determined as indicating motion by visual assessment of the image sequence. Relative to this standard of reference, the software reached 80% sensitivity versus 67% specificity.

Conclusions: Optical flow seems to be an efficient concept for automated detection of patient motion on the projection images acquired during a CBCT scan.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.oooo.2014.12.008DOI Listing

Publication Analysis

Top Keywords

projection images
16
automated detection
12
detection patient
12
cbct scan
12
patient motion
12
patient movement
8
motion projection
8
images acquired
8
optical flow
8
visual assessment
8

Similar Publications

Ossification of the ligamentum flavum (OLF) can lead to dural ossification, significantly increasing the risk of complications, including intraoperative nerve injury. The application of augmented reality (AR) and advanced digital technologies in spine surgery has the potential to reduce these risks. This case report highlights a perioperative nerve injury-free microsurgery using elastic image fusion technology, which integrates preoperative imaging with intraoperative computed tomography for a patient with severe stenotic OLF and dural ossification.

View Article and Find Full Text PDF

The study aim was to evaluate whether reducing bed position acquisition time would result in significant detriment to image quality. Secondary aims were to compare effect of time of flight (TOF) and Q.Clear reconstructions and patient BMI on image quality.

View Article and Find Full Text PDF

In targeted alpha-particle therapy, actinium-225 (Ac-225) has emerged as a radionuclide of potential, driving extensive efforts to develop innovative radiopharmaceuticals. High-resolution imaging of alpha particles is required for precisely detecting alpha-emitting radionuclides in cellular environments and small organs. Here, we report real-time trajectory imaging of alpha particles emitted by Ac-225 and its daughter radionuclides, utilizing an alpha particle trajectory imaging system.

View Article and Find Full Text PDF

When we collaborate with others to tackle novel problems, we anticipate how they will perform their part of the task to coordinate behavior effectively. We might estimate how well someone else will perform by extrapolating from estimates of how well we ourselves would perform. This account predicts that our metacognitive model should make accurate predictions when projected onto people as good as, or worse than, us but not on those whose abilities exceed our own.

View Article and Find Full Text PDF

Recent advances in microCT are facilitating the investigation of microstructures in spiders and insects leading to an increased number of studies investigating their neuroanatomy. Although microCT is a powerful tool, its effectiveness depends on appropriate tissue preparation and scan settings, particularly for soft, non-sclerotized tissues, such as muscles, organs, and neural tissues. As the application of microCT in spiders is only in its infancy, published protocols are often difficult to implement due to substantial size variation of the specimens.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!