Background: The benefits of implantable cardioverter-defibrillators (ICDs) have been well demonstrated in many clinical trials, and ICD shocks for ventricular tachyarrhythmias save lives. However, inappropriate and unnecessary shock delivery remains a significant clinical issue with considerable consequences for patients and the healthcare system.

Objective: The purpose of the PainFree SmartShock Technology (SST) study was to investigate new-generation ICDs to reduce inappropriate and unnecessary shocks through novel discrimination algorithms with modern programming strategies.

Methods: This prospective, multicenter clinical trial enrolled 2790 patients with approved indication for ICD implantation (79% male, mean age 65 years; 69% primary prevention indication, 27% single-chamber ICD, 33% replacement or upgrade). Patients were followed for a minimum of 12 months, and mean follow-up was 22 months. The primary end-point of the study was the percentage of patients remaining free of inappropriate shocks at 1 year postimplant, analyzed separately for dual/triple-chamber ICDs (N = 2019) and single-chamber ICDs (N = 751).

Results: The inappropriate shock rate at 1 year was 1.5% for patients with dual/triple-chamber ICDs and 2.5% for patients with single-chamber devices. Two years postimplant, the inappropriate shock rate was 2.8% for patients with dual-/triple chamber ICDs and 3.7% for those with single-chamber ICDs. The most common cause of an inappropriate shock in both groups was atrial fibrillation or flutter.

Conclusion: In a large patient cohort receiving ICDs for primary or secondary prevention, the adoption of novel enhanced detection algorithms in conjunction with routine implementation of modern programming strategies led to a very low inappropriate shock rate.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.hrthm.2015.01.017DOI Listing

Publication Analysis

Top Keywords

inappropriate shock
20
shock rate
12
low inappropriate
8
patients
8
implantable cardioverter-defibrillators
8
detection algorithms
8
icds
8
inappropriate unnecessary
8
modern programming
8
dual/triple-chamber icds
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!