Hemodynamic changes and retrograde flow in LVAD failure.

ASAIO J

From the *Department of Bioengineering, Cardiovascular Innovation Institute, University of Louisville, Louisville, Kentucky 40292; †Department of Cardiovascular and Thoracic Surgery, Cardiovascular Innovation Institute, University of Louisville, Louisville, Kentucky 40202; ‡Eidgenössische Technische Hochschule Hönggerberg, Zürich, Switzerland; and §Division of Cardiovascular Surgery, Hospital of the University of Pennsylvania, Philadelphia, Pennsylvania 19104.

Published: February 2016

In the event of left ventricular assist device (LVAD) failure, we hypothesized that rotary blood pumps will experience significant retrograde flow and induce adverse physiologic responses. Catastrophic LVAD failure was investigated in computer simulation with pulsatile, axial, and centrifugal LVAD, mock flow loop with pulsatile (PVAD) and centrifugal (ROTAFLOW), and healthy and chronic ischemic heart failure bovine models with pulsatile (PVAD), axial (HeartMate II), and centrifugal (HVAD) pumps. Simulated conditions were LVAD "off" with outflow graft clamped (baseline), LVAD "off" with outflow graft unclamped (LVAD failure), and LVAD "on" (5 L/min). Hemodynamics (aortic and ventricular blood pressures, LVAD flow, and left ventricular volume), echocardiography (cardiac volumes), and end-organ perfusion (regional blood flow microspheres) were measured and analyzed. Retrograde flow was observed with axial and centrifugal rotary pumps during LVAD failure in computer simulation (axial = -3.4 L/min, centrifugal = -2.8 L/min), mock circulation (pulsatile = -0.1 L/min, centrifugal = -2.7 L/min), healthy (pulsatile = -1.2 ± 0.3 L/min, axial = -2.2 ± 0.2 L/min, centrifugal = -1.9 ± 0.3 L/min), and ischemic heart failure (centrifugal = 2.2 ± 0.7 L/min) bovine models for all test conditions (p < 0.05). Differences between axial and centrifugal LVAD were statistically indiscernible. Retrograde flow increased ventricular end-systolic and end-diastolic volumes and workload, and decreased myocardial and end-organ perfusion during LVAD failure compared with baseline, LVAD support, and pulsatile LVAD failure.

Download full-text PDF

Source
http://dx.doi.org/10.1097/MAT.0000000000000200DOI Listing

Publication Analysis

Top Keywords

lvad failure
28
retrograde flow
16
lvad
14
axial centrifugal
12
l/min centrifugal
12
failure
9
centrifugal
9
l/min
9
left ventricular
8
computer simulation
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!