Different chemiluminescence systems based on luminol, permanganate, manganese(IV) and cerium(IV) reagents were compared regarding their sensitivity and selectivity to determine plant polyphenols. Among the seventeen systems tested, Mn(IV)-formaldehyde-hexametaphosphate was considered to be the most suitable for polyphenols detection. The developed flow injection method (FI-CL) based on enhancing effect of polyphenols on Mn(IV) chemiluminescence is characterised by low detection limit of gallic acid (0.02μgL(-1)) and high precision (RSD=1.7%). The calibration graph was linear from 0.1 to 100μgL(-1). The selectivity studies revealed that the FI-CL method ensures accurate determination of the total polyphenols content in food samples. The method was successfully applied to analysis of a variety of plant-derived foods (wine, tea, cereal coffee, fruit and vegetable juices, herbs and spices). The proposed method is superior to conventional spectrophotometric assays due to its higher sample throughput (195samplesh(-1)), simplicity, sensitivity and, above all, higher selectivity.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1016/j.foodchem.2014.12.053 | DOI Listing |
Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!