Objective: An epidemic of chronic kidney disease (CKD) has been identified in Pacific coastal regions of Central America, and screening in the field in these low income countries remains logistically problematic. We tested the performance characteristics of a point of care creatinine analyzer compared to standardized serum creatinine measurements.

Methods: Measurements were conducted in 100 persons from a local health center (n=34) and hospital (n=66) in Rivas, Nicaragua using both a point-of-care analyzer (StatSensor Xpress, Nova Biomedical) and serum creatinine by Jaffe kinetic method with a Roche Cobas Integra 400 analyzer. Percent coefficient of variation, sensitivity and specificity of the StatSensor Xpress were determined.

Results: The average coefficient of variation (CV) was 1.28% for the serum creatinine and CV for the StatSensor Xpress analyzer was 6.8%. The median intra-individual creatinine results obtained with the StatSensor Xpress device were 0.32 mg/dL higher than those by serum creatinine by Jaffe kinetic method. The sensitivity and specificity of the StatSensor Xpress device for identifying subjects with abnormal creatinine (defined as >1.2 mg/dL) was 100% and 79%, respectively.

Conclusions: Point of care testing for creatinine demonstrated acceptable repeatability, excellent sensitivity (100%) and modest specificity (79%). Using the point of care testing will allow for generalized screening in the field in low income countries; however, confirmation for elevated levels >1.2 mg/dL will require a second laboratory test confirmation.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.clinbiochem.2015.01.005DOI Listing

Publication Analysis

Top Keywords

statsensor xpress
20
serum creatinine
16
point care
12
creatinine
9
chronic kidney
8
kidney disease
8
rivas nicaragua
8
screening field
8
field low
8
low income
8

Similar Publications

Clinical application of the StatSensor and StatSensor Xpress point-of-care creatinine measurement devices in dogs.

Vet Clin Pathol

December 2022

Department of Clinical Sciences and Advanced Medicine, University of Pennsylvania, School of Veterinary Medicine, Philadelphia, Pennsylvania, USA.

Article Synopsis
  • Creatinine is a key blood marker for diagnosing and monitoring kidney issues, but there are challenges in getting accurate measurements at the bedside for both humans and pets.
  • This study tested the accuracy and precision of two handheld creatinine measuring devices (StatSensor and StatSensor Xpress) in dogs with varying levels of kidney function.
  • Results showed that both devices had significant errors and variances, indicating they are not reliable for clinical use, highlighting the need for better point-of-care creatinine testing options in veterinary medicine.
View Article and Find Full Text PDF

Background: Patients with low estimated glomerular filtration rates may be at higher risk of post-contrast acute kidney injury following contrast-enhanced computed tomography imaging. Point-of-care devices allow rapid measurement of estimated glomerular filtration rates for patients referred without a recent estimated glomerular filtration rate result.

Objectives: To assess the clinical effectiveness and cost-effectiveness of point-of-care creatinine tests for outpatients without a recent estimated glomerular filtration rate measurement who need contrast-enhanced computed tomography imaging.

View Article and Find Full Text PDF

Background: The high frequency of outpatient visits after kidney transplantation is burdensome to both the recovering patient and health care capacity. Self-monitoring kidney function offers a promising strategy to reduce the number of these outpatient visits.

Objective: The objective of this study was to investigate whether it is safe to rely on patients' self-measurements of creatinine and blood pressure, using data from a self-management randomized controlled trial.

View Article and Find Full Text PDF

Objectives The aim of the study was to evaluate whether a handheld creatinine analyser (StatSensor Xpress; SSXp), available for human patients, can be used to measure creatinine reliably in cats. Methods Analytical performance was evaluated by determining within- and between-run coefficient of variation (CV, %), total error observed (TE, %) and sigma metrics. Fifty client-owned cats presenting for investigation of clinical disease had creatinine measured simultaneously, using SSXp (whole blood and plasma) and a reference instrument (Konelab, serum); 48 paired samples were included in the study.

View Article and Find Full Text PDF

Creatinine is a parameter that is required to monitor renal function and is important to follow in patients under treatment with potentially toxic renal drugs, such as the anti-HIV drug Tenofovir. A point of care instrument to measure creatinine would be useful for patients monitoring in resource-limited settings, where more instruments that are sophisticated are not available. The StatSensor Xpress Creatinine (Nova Biomedical Cooperation, Waltham, MA, USA) point of care analyzer was evaluated for its diagnostic performance in indicating drug therapy change.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!