Unlabelled: A major disadvantage of (18)F-FDG PET involves poor detection of small lesions and lesions with low metabolism, caused by limited spatial resolution and relatively large image voxel size. As spatial resolution and sensitivity are better in new PET systems, it is expected that small-lesion detection could be improved using smaller voxels. The aim of this study was to test this hypothesis using a state-of-the-art time-of-flight PET/CT device.

Methods: (18)F-FDG PET scans of 2 image-quality phantoms (sphere sizes, 4-37 mm) and 39 consecutive patients with lung cancer were analyzed on a time-of-flight PET/CT system. Images were iteratively reconstructed with standard 4 × 4 × 4 mm voxels and smaller 2 × 2 × 2 mm voxels. For the phantom study, we determined contrast-recovery coefficients and signal-to-noise ratios (SNRs). For the patient study, (18)F-FDG PET-positive lesions in the chest and upper abdomen with a volume less than 3.0 mL (diameter, <18 mm) were included. Lesion mean and maximum standardized uptake values (SUVmean and SUVmax, respectively) were determined in both image sets. SNRs were determined by comparing SUVmax and SUVmean with background noise levels. A subanalysis was performed for lesions less than 0.75 mL (diameter, <11 mm). For qualitative analysis of patient data, 3 experienced nuclear medicine physicians gave their preference after visual side-by-side analysis.

Results: For phantom spheres 13 mm or less, we found higher contrast-recovery coefficients and SNRs using small-voxel reconstructions. For 66 included (18)F-FDG PET-positive lesions, the average increase in SUVmean and SUVmax using the small-voxel images was 17% and 32%, respectively (P < 0.01). For lesions less than 0.75 mL (21 in total), the average increase was 21% and 44%, respectively. Moreover, averaged over all lesions, the mean and maximum SNR increased by 20% and 27%, respectively (P < 0.01). For lesions less than 0.75 mL, these values increased up to 23% and 46%, respectively. The physicians preferred the small-voxel reconstructions in 76% of cases.

Conclusion: Supported by a phantom study, there was a visual preference toward (18)F-FDG PET images reconstructed with 2 × 2 × 2 mm voxels and a profound increase in standardized uptake value and SNR for small lesions. Hence, it is expected that small-lesion detection improves using small-voxel reconstructions.

Download full-text PDF

Source
http://dx.doi.org/10.2967/jnmt.114.147215DOI Listing

Publication Analysis

Top Keywords

time-of-flight pet/ct
12
detection small
8
small lesions
8
state-of-the-art time-of-flight
8
pet/ct system
8
18f-fdg pet
8
spatial resolution
8
smaller voxels
8
improving detection
4
lesions
4

Similar Publications

Optimizing time-of-flight of PET/CT image quality via penalty β value in Bayesian penalized likelihood reconstruction algorithm.

Radiography (Lond)

December 2024

Department of Physics, Faculty of Science, Universiti Putra Malaysia, 43400 Serdang, Selangor, Malaysia. Electronic address:

Introduction: Optimizing the image quality of Positron Emission Tomography/Computed Tomography (PET/CT) systems is crucial for effective monitoring, diagnosis, and treatment planning in oncology. This study evaluates the impact of time-of-flight (TOF) on PET/CT performance, focusing on varying penalty β values within Q. Clear reconstruction algorithm.

View Article and Find Full Text PDF

Background: Treatment planning in radiation therapy (RT) is performed on image sets acquired with commercial x-ray computed tomography (CT) scanners. Considering an increased frequency of verification scans for adaptive RT and the advent of alternatives to x-ray CTs, there is a need to review the requirements for image sets used in RT planning.

Purpose: This study aims to derive the required image quality (IQ) for the computation of the dose distribution in proton therapy (PT) regarding spatial resolution and the combination of spatial resolution and noise.

View Article and Find Full Text PDF

: The aim is to propose a model for predicting occult paraaortic lymph node (PALN) involvement in locally advanced cervical cancer (LACC) patients by including parameters such as reconstruction detection technology (use of time-of-flight) and parameters related to the primary tumor. This model will then be compared with the scores used in routine clinical practice; : This retrospective observational cohort study included patients diagnosed with LACC who underwent F-FDG PET/CT prior to PALN surgical staging between February 2012 and May 2020. The following parameters were collected on PET/CT: tumor SUVmax, tumor MTV, number of common and distal pelvic node involvements.

View Article and Find Full Text PDF

Feasibility of PET-enabled dual-energy CT imaging: First physical phantom and initial patient study results.

Eur J Nucl Med Mol Imaging

November 2024

Department of Radiology, UC Davis Health, 95817, Sacramento, CA, USA.

Purpose: Dual-energy (DE) CT enables material decomposition by using two different x-ray energies and may be combined with PET for improved multimodality imaging. However, this increases radiation dose and may require a hardware upgrade due to the added second x-ray CT scan. The recently proposed PET-enabled DECT method allows dual-energy imaging using a conventional PET/CT scanner without the need to change scanner hardware or increase radiation exposure.

View Article and Find Full Text PDF

Performance Characteristics of a New Generation 148-cm Axial Field-of-View uMI Panorama GS PET/CT System with Extended NEMA NU 2-2018 and EARL Standards.

J Nucl Med

December 2024

Department of Nuclear Medicine, State Key Laboratory of Complex Severe and Rare Diseases, Beijing Key Laboratory of Molecular Targeted Diagnosis and Therapy in Nuclear Medicine, Peking Union Medical College Hospital, Chinese Academy of Medical Sciences, Beijing, China

Article Synopsis
  • The uMI Panorama GS PET/CT system is a new advanced imaging scanner designed to provide high sensitivity and excellent image resolution for medical diagnosis.
  • The study evaluates the system's performance based on established standards, focusing on parameters like spatial resolution, count rate performance, and sensitivity, using various testing methodologies.
  • Results showed impressive spatial resolutions below 3mm, high sensitivity rates, and good image quality with compliance to both NEMA and EARL guidelines, indicating the system's effectiveness for clinical use.
View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!