A robust adaptive k-space sampling method is proposed for faster acquisition and reconstruction of MR images. In this method, undersampling patterns are generated based on magnitude profile of a fully acquired 2-D k-space data. Images are reconstructed using compressive sampling reconstruction algorithm. Simulation experiments are done to assess the performance of the proposed method under various signal-to-noise ratio (SNR) levels. The performance of the method is better than non-adaptive variable density sampling method when k-space SNR is greater than 10dB. The method is implemented on a fully acquired multi-slice raw k-space data and a quality assurance phantom data. Data reduction of up to 60% is achieved in the multi-slice imaging data and 75% is achieved in the phantom imaging data. The results show that reconstruction accuracy is improved over non-adaptive or conventional variable density sampling method. The proposed sampling method is signal dependent and the estimation of sampling locations is robust to noise. As a result, it eliminates the necessity of mathematical model and parameter tuning to compute k-space sampling patterns as required in non-adaptive sampling methods.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1016/j.mri.2015.01.008 | DOI Listing |
Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!