Severity: Warning
Message: file_get_contents(https://...@pubfacts.com&api_key=b8daa3ad693db53b1410957c26c9a51b4908&a=1): Failed to open stream: HTTP request failed! HTTP/1.1 429 Too Many Requests
Filename: helpers/my_audit_helper.php
Line Number: 176
Backtrace:
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 176
Function: file_get_contents
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 250
Function: simplexml_load_file_from_url
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 3122
Function: getPubMedXML
File: /var/www/html/application/controllers/Detail.php
Line: 575
Function: pubMedSearch_Global
File: /var/www/html/application/controllers/Detail.php
Line: 489
Function: pubMedGetRelatedKeyword
File: /var/www/html/index.php
Line: 316
Function: require_once
Background And Objectives: The use of pulse-oximetry screening to detect critical congenital heart defects in newborns has gained national and international momentum in the past decade. Our hospital system began screening in 2008. Since then, our program has undergone leadership changes and multiple quality improvement interventions. The aims of this study are to evaluate the evolution of our pulse-oximetry program and to provide insights from lessons learned over the course of a long-standing program.
Methods: We reviewed 6 years of screening data and evaluated trends of missed screens, false-positives, protocol violations, and parental decline of screening. We implemented 3 quality improvement interventions (change in protocol, redesign of an electronic medical record documentation system to autocalculate results, and transition from research to standard-of-care) and reviewed the impact of a rigorous quality assurance review process. We used linear regression and statistical process control charts to evaluate the data.
Results: A total of 18,363 newborns were screened; we identified 5 critical cases. We observed a significant decrease in missed (P < .001) and false-positive (P = .03) screens over time but found no significant trend in the rate of percentage of protocol violations (P = .26) or decline of screening (P = .99). Each metric showed behavior attributable to at least 1 quality improvement intervention.
Conclusions: We established a sustainable pulse-oximetry screening program in our community hospital system, and the screening has now become routine. The quality of our screening was influenced by choice of screening protocol, rigor of quality assurance reviews, and the process used to interpret screening results.