With millions of at-risk people undiagnosed with pre-diabetes and diabetes, there is a need to identify alternate screening sites for out-of-range glucose values. We examined practical issues and accuracy (relative to High Performance Liquid Chromatography testing in a laboratory) in the use of the A1cNow point of care device for this screening in general practice dental clinics at a large University-based Dental College. Health care professionals obtained evaluable readings for only 70% of the subjects, even after two attempts, and its use according to manufacturer's instructions was often challenging in the busy environment of the dental clinic. At thresholds for pre-diabetes and diabetes established by the American Diabetes Association, sensitivities of the A1cNow kit relative to the HPLC method were 91.9% and 100%, respectively. However, specificities for pre-diabetes and diabetes were 66.7% and 82.4%, respectively, indicating many false positive results. A better strategy for diabetes screening may involve a laboratory-based analysis approach that is patient- and provider-friendly, with minimal burden to the dental team.