The effects of voxel size on cone beam computed tomography images of the temporomandibular joints.

Oral Surg Oral Med Oral Pathol Oral Radiol

Professor and the Dr. Lloyd & Mrs. Kay Chapman Chair in Clinical Sciences, Discipline of Oral and Maxillofacial Radiology, Faculty of Dentistry, the University of Toronto, Toronto, ON, Canada. Electronic address:

Published: February 2015

Objectives: This clinical study assesses the effect of cone beam computed tomography (CBCT) voxel size on the ability to detect osseous changes associated with degenerative disease of the temporomandibular joint (TMJ). The effect of voxel size on perceived CBCT image quality is also evaluated.

Study Design: Twenty-two patients presenting for TMJ imaging with suspected degenerative disease were imaged with the Carestream 9000 CBCT unit, using separate right and left joint acquisitions (n = 44). Images were archived at native and downsampled voxel resolutions of 76 μm and 300 μm, respectively. Three oral and maxillofacial radiologists evaluated the images for osseous changes, as well as image quality by using a visual analog scale.

Results: There was no statistically significant difference between the voxel sizes in the detection of TMJ osteoarthritic changes. The mean visual analog scale response did, however, differ significantly between the two groups (P = .02).

Conclusions: Despite no improvement in diagnostic efficacy with a smaller voxel size, perceived image quality is consistently higher for images with greater spatial resolution.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.oooo.2014.10.015DOI Listing

Publication Analysis

Top Keywords

voxel size
16
image quality
12
cone beam
8
beam computed
8
computed tomography
8
osseous changes
8
degenerative disease
8
size perceived
8
visual analog
8
voxel
5

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!