Background: The outcome of cementless total hip arthroplasty depends on many factors. We must not forget fundamental things those are design of outer surface of the component, that leads bone ingrowth into the prosthesis, better initial stability, and better insertional techniques. The purpose of this study was to review our experience with metal-on-metal total hip arthroplasty with a Wagner standard cup for patients who had acetabular dysplasia.

Patients And Methods: Fifty-four patients with 55 hips underwent primary metal-on-metal total hip arthroplasty (Metasul prosthesis) with a Wagner standard cup (44-48 mm in outer diameter) and were followed for a minimum of 10 years. All patients received the same type of cementless femoral component (Natural hip stem) and femoral head (28 mm in diameter).

Results: Seventeen of the 55 Wagner standard cups (30.9%) showed aseptic loosening over a mean period of 3.6 years after surgery, and there were no bone anchors on the outer surface of the 16 retrieved cups.

Conclusion: From our experience, the small Wagner standard cup does not achieve sufficient osteointegration and we do not recommend the use of this cup, especially for patients with acetabular dysplasia and/or those with a small stature.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC4281638PMC
http://dx.doi.org/10.1016/j.jor.2014.08.003DOI Listing

Publication Analysis

Top Keywords

wagner standard
20
total hip
16
hip arthroplasty
16
standard cup
16
arthroplasty wagner
8
outer surface
8
metal-on-metal total
8
cup patients
8
patients acetabular
8
hip
6

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!