Intravascular optical coherence tomography light scattering artifacts: merry-go-rounding, blooming, and ghost struts.

J Biomed Opt

The University of Texas Health Science Center at San Antonio, Division of Cardiology, Department of Medicine, 7703 Floyd Curl Drive, San Antonio, Texas 78229, United StatesbSouth Texas Veterans Health Care System, The Department of Veterans Affairs, San A.

Published: December 2014

We sought to elucidate the mechanisms underlying two common intravascular optical coherence tomography (IV-OCT) artifacts that occur when imaging metallic stents: “merry-go-rounding” (MGR), which is an increase in strut arc length (SAL), and “blooming,” which is an increase in the strut reflection thickness (blooming thickness). Due to uncontrollable variables that occur in vivo, we performed an in vitro assessment of MGR and blooming in stented vessel phantoms. Using Xience V and Driver stents, we examined the effects of catheter offset, intimal strut coverage, and residual blood on SAL and blooming thickness in IV-OCT images. Catheter offset and strut coverage both caused minor MGR, while the greatest MGR effect resulted from light scattering by residual blood in the vessel lumen, with 1% hematocrit (Hct) causing a more than fourfold increase in SAL compared with saline (p<0.001 ). Residual blood also resulted in blooming, with blooming thickness more than doubling when imaged in 0.5% Hct compared with saline (p<0.001 ). We demonstrate that a previously undescribed mechanism, light scattering by residual blood in the imaging field, is the predominant cause of MGR. Light scattering also results in blooming, and a newly described artifact, three-dimensional-MGR, which results in “ghost struts” in B-scans.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC4659478PMC
http://dx.doi.org/10.1117/1.JBO.19.12.126017DOI Listing

Publication Analysis

Top Keywords

intravascular optical
8
optical coherence
8
coherence tomography
8
light scattering
8
increase strut
8
blooming thickness
8
catheter offset
8
strut coverage
8
residual blood
8
tomography light
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!