Purpose: Dental professionals suffer from a high prevalence of work-related musculoskeletal disorders (MSD). Dental hygienists in particular have a high prevalence of pain in the forearms and hands. The objective of this study was to compare 1 cordless handpiece to 2 corded handpieces during simulated tooth polishing in terms of the muscle loads (recorded as electromyography (EMG) activity), duration of polishing procedure, and dental hygienist opinion about ease of use.

Methods: EMG was used to quantify muscle electrical activity of 4 forearm muscles during simulated dental polishing with 2 corded handpieces (HP-A and HP-B) and 1 cordless handpiece (HP-C). A convenience sample of 30 dental hygienists (23 to 57 years of age) with 1 to 20+ years of clinical practice experience completed the study. Each participant spent approximately 5 minutes polishing 3 predetermined teeth in each of the 4 quadrants. The sequence of the handpieces was randomly assigned. At the end of the study, participants completed a subjective end user evaluation of handpiece preference.

Results: Muscle activity levels of 10th, 50th and 90th percentiles did not differ significantly between the 3 handpieces tested (p>0.05). However, total muscle workload (integrated EMG) was lowest for the cordless handpiece (HP-C), but this was only significantly less than HP-A (p<0.05). Polishing using the cordless handpiece (HP-C) (M=257 seconds, SD=112 seconds) took significantly less time than either the HP-A corded (M=290 seconds, SD=137 seconds) or HP-B corded handpiece (M=290 seconds, SD=126 seconds) (p<0.05). Overall, 50% of the study participants preferred the cordless handpiece, 37% preferred HP-A and 13% preferred HP-B (p<0.05).

Conclusion: Use of the cordless handpiece reduced the duration of polishing, which in turn led to less total muscle activity, but not muscle intensity. Overall, dental hygienists preferred the cordless handpiece.

Download full-text PDF

Source

Publication Analysis

Top Keywords

cordless handpiece
12
muscle activity
8
simulated tooth
8
tooth polishing
8
high prevalence
8
dental hygienists
8
corded handpieces
8
handpiece hp-c
8
handpieces
5
muscle
5

Similar Publications

Introduction: This ex vivo study evaluated the accuracy of the Electronic Apex Locator (EAL) and Automatic Apical Stop (AAS) functions of the E-Connect S+ and Morita Tri Auto ZX2+ cordless apex locators in determining patency length.

Methods: Sixty-four human teeth with a single root were randomly allocated into E-connect or Morita groups (n = 32). The canals were accessed and preflared, after which a size 15 K-file was inserted into the canal to the major foramen and recorded as the actual length (AL).

View Article and Find Full Text PDF

Purpose: Dental professionals suffer from a high prevalence of work-related musculoskeletal disorders (MSD). Dental hygienists in particular have a high prevalence of pain in the forearms and hands. The objective of this study was to compare 1 cordless handpiece to 2 corded handpieces during simulated tooth polishing in terms of the muscle loads (recorded as electromyography (EMG) activity), duration of polishing procedure, and dental hygienist opinion about ease of use.

View Article and Find Full Text PDF

Objective: On April 25, 2012, the first laparoscopic cordless ultrasonic device (Sonicision, Covidien, Mansfield, Massachusetts) was used in a clinical setting. We describe our initial experience.

Methods: The cordless device is assembled with a reusable battery and generator on a base hand-piece.

View Article and Find Full Text PDF

Aim: The aim of this study was to assess the clinical perfomance of a cordless handpiece with a built-in apex locator - the Tri Auto ZX - designed for root canal preparation with nickel-titanium rotary files.

Methodology: Twenty-five human maxillary incisor and canine teeth scheduled for extraction with mature apices were selected for the study. Informed written consent was obtained from each patient before treatment.

View Article and Find Full Text PDF

Apical extent of rotary canal instrumentation with an apex-locating handpiece in vitro.

Oral Surg Oral Med Oral Pathol Oral Radiol Endod

March 1998

Department of Endodontics, Faculty of Dentistry, University of Toronto, Canada.

Problem: The Tri Auto ZX (J. Morita Co., Kyoto, Japan) is a cordless endodontic handpiece with a built-in apex locator that is programmed to reverse the direction of rotation when the file reaches a predetermined apical level or when torque becomes excessive.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!