Background: The interference screw technique is commonly used in tenodesis of the long head of the biceps for its well-documented robust fixation strength. Some complications may occur after tenodesis with interference screw such as persistent pain, bone fracture and cyst formation. A new technique using a small "soft anchor" has been proposed to avoid the risk of occurrence of the above-mentioned complications associated with the use of the interference screw. However, the proposed technique must provide adequate fixation strength. This study investigated the mechanical performance of the new technique and compared it with interference screw fixation.

Methods: Fourteen human humeri and proximal biceps were tested after tenodesis using the two techniques. The fixation constructs were cycled 500 times between 20N and 100N at 1Hz to simulate some level of post-operative physical activity. Then, a tensile test to failure was performed to determine the strength of the two tenodesis constructs.

Findings: The ultimate strength was 238N (SD 96N) and 172N (SD 58N) for the "soft anchor" and the interference screw, respectively (P=0.14). In two out of seven repetitions in both groups, failure occurred at low load level due to inaccuracies in performing tenodesis. Considering these cases as outliers, the strength values increased up to 290N (SD 40N) and 202N (SD 32N) for the "soft anchor" and the interference screw, respectively (P=0.02).

Interpretation: The "soft anchor" technique provides a fixation strength comparable with the interference screw, but without using a screw. It could be considered as an alternative for suprapectoral biceps tenodesis.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.clinbiomech.2014.12.001DOI Listing

Publication Analysis

Top Keywords

interference screw
32
"soft anchor"
20
fixation strength
12
screw
9
suprapectoral biceps
8
tenodesis
8
biceps tenodesis
8
interference
8
anchor" interference
8
technique
6

Similar Publications

Background: Surgical techniques for biceps tenodesis vary in approach, fixation strategy, and anatomic location without clear superior technique for this common procedure.

Hypothesis/purpose: The purpose of this study was to prospectively evaluate a randomized cohort of patients undergoing arthroscopic suprapectoral (ASBT) with interference screw fixation using an inlay technique versus mini-open subpectoral (MOBT) with a unicortical button implant using an onlay technique with regards to 1) clinical outcome measures and 2) structural healing as evaluated by ultrasound.

Methods: From May 2017 to April 2021, patients undergoing biceps tenodesis were preoperatively randomized to either ASBT or MOBT.

View Article and Find Full Text PDF

Anatomic Distal Biceps Tendon Repair With All-Suture Cortical Buttons.

Arthrosc Tech

December 2024

Department of Orthopaedic Surgery, University of California Irvine, Orange, California, U.S.A.

Acute, traumatic distal biceps tendon ruptures are a common injury in the middle-aged athletic male population, with direct anatomic surgical repair being the most effective technique to restore maximal strength. Multiple techniques for distal biceps tendon repair have been described, including single- or dual-incision approaches and tendon fixation with cortical buttons, interference screws, suture anchors, and transosseous sutures. In this Technical Note, we demonstrate an anatomic distal biceps tendon repair technique with a single-incision approach using 2 all-suture cortical buttons.

View Article and Find Full Text PDF

Background: An all-inside endoscopic flexor hallucis longus (FHL) tendon transfer is indicated for the treatment of chronic, full-thickness Achilles tendon defects. The aim of this procedure is to restore function of the gastrocnemius-soleus complex while avoiding the wound complications associated with open procedures.

Description: This procedure can be performed through 2 endoscopic portals, a posteromedial portal (the working portal) and a posterolateral portal (the visualization portal).

View Article and Find Full Text PDF

Background: A lateral extra-articular tenodesis (LET) is increasingly being utilized to augment an anterior cruciate ligament reconstruction because it has been shown to reduce the risk of postreconstruction graft failure or recurrent rotatory instability. Various femoral fixation techniques are available, including the use of an interference screw, staple, or suture anchor.

Purpose: To determine and compare the biomechanical properties of an LET graft when using an interference screw, staple, or suture anchor for the femoral fixation for a modified Lemaire LET.

View Article and Find Full Text PDF

Purpose: To biomechanically evaluate a flat posterior cruciate ligament (PCL) reconstruction utilizing rectangular femoral bone tunnels.

Methods: Eight fresh-frozen human knee specimens were tested in a six-degrees-of-freedom robotic test setup. In each testing step, a force-controlled test protocol was performed, including 89 N posterior tibial translation (PTT) in neutral, internal and external rotation, from 0 to 90° of flexion.

View Article and Find Full Text PDF