Introduction: High-yielding, automated production of a PET tracer that reflects proliferation, 3'-deoxy-3'-[(18)F]fluorothymidine ([(18)F]FLT), is reported using a modified Bioscan Coincidence FDG reaction module.
Methods: Production of [(18)F]FLT was implemented through: (1) modification of an original FDG manifold; (2) application of an alternate time sequence; and (3) altered solid-phase extraction (SPE) purification. Quality control testing, including standard radiochemical figures of merit and preclinical positron emission tomography (PET) imaging, was carried out.
Results: High decay-corrected yields of [(18)F]FLT (16-39%) were reproducibly obtained. The product exhibited very high specific activity (4586.9TBq/mmol; 123,969Ci/mmol) and radiochemical purity (>99%). Overall, the [(18)F]FLT produced in this manner was superior to typical productions that utilized a GE TRACERlab FXF-N reaction module. Additionally, purification with SPE cartridges, followed by manual elution, accelerated overall run time and resulted in a two-fold increase in [(18)F]FLT concentration. PET imaging showed the [(18)F]FLT produced by this method was highly suitable for non-invasive tumor imaging in mice.
Conclusions: The Bioscan Coincidence GE FDG Reaction Module was readily adapted to reproducibly provide [(18)F]FLT in high yield, specific activity, and radiochemical purity. The approach was suitable to provide sufficient amounts of material for preclinical studies.
Download full-text PDF |
Source |
---|---|
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC4867542 | PMC |
http://dx.doi.org/10.1016/j.apradiso.2014.11.011 | DOI Listing |
Appl Radiat Isot
March 2015
Vanderbilt University Institute of Imaging Science (VUIIS), Vanderbilt University Medical Center, Nashville, TN, USA; Program in Chemical and Physical Biology, Vanderbilt University Medical Center, Nashville, TN, USA; Department of Biomedical Engineering, Vanderbilt University, Nashville, TN, USA; Department of Radiology and Radiological Sciences, Vanderbilt University Medical Center, Nashville, TN, USA; Vanderbilt-Ingram Cancer Center (VICC), Vanderbilt University Medical Center, Nashville, TN, USA; Department of Neurosurgery, Vanderbilt University Medical Center, Nashville, TN 37232, USA. Electronic address:
Introduction: High-yielding, automated production of a PET tracer that reflects proliferation, 3'-deoxy-3'-[(18)F]fluorothymidine ([(18)F]FLT), is reported using a modified Bioscan Coincidence FDG reaction module.
Methods: Production of [(18)F]FLT was implemented through: (1) modification of an original FDG manifold; (2) application of an alternate time sequence; and (3) altered solid-phase extraction (SPE) purification. Quality control testing, including standard radiochemical figures of merit and preclinical positron emission tomography (PET) imaging, was carried out.
J Nucl Med
March 1998
Center for Positron Emission Tomography, Emory University, Atlanta, Georgia 30322, USA.
Unlabelled: This study evaluated the Pico-Count (Bioscan, Inc., Washington, DC) flow-through radioactivity detector, designed for use in PET studies of cerebral blood flow.
Methods: The Pico-Count detects the two 511-keV positron annihilation photons with two bismuth germanate detectors operating in coincidence.
Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!