The objective of this study was to analyze the influence of the increment of gantry angle and the number of arcs on esophageal volumetric modulated arc therapy plan. All plans were done in Monaco planning system for Elekta Synergy linear accelerator with 80 multileaf collimator (MLC). Volumetric modulated arc therapy (VMAT) plans were done with different increment of gantry angle like 15°, 20°, 30° and 40°. The remaining parameters were similar for all the plans. The results were compared. To compare the plan quality with number of arcs, VMAT plans were done with single and dual arc with increment of gantry angle of 20°. The dose to gross tumor volume (GTV) for 60 Gy and planning target volume (PTV) for 48 Gy was compared. The dosimetric parameters D98%, D95%, D50% and Dmax of GTV were analyzed. The homogeneity index (HI) and conformity index (CI) of GTV were studied and the dose to 98% and 95% of PTV was analyzed. Maximum dose to spinal cord and planning risk volume of cord (PRV cord) was compared. The Volume of lung receiving 10 Gy, 20 Gy and mean dose was analyzed. The volume of heart receiving 30 Gy and 45 Gy was compared. The volume of normal tissue receiving greater than 2 Gy and 5 Gy was compared. The number of monitor units (MU) required to deliver the plans were compared. The plan with larger increment of gantry angle proved to be superior to smaller increment of gantry angle plans in terms of dose coverage, HI, CI and normal tissue sparing. The number of arcs did not make any difference in the quality of the plan.
Download full-text PDF |
Source |
---|---|
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC4258731 | PMC |
http://dx.doi.org/10.4103/0971-6203.144488 | DOI Listing |
J Appl Clin Med Phys
November 2024
Department of Radiation Oncology, Chung-Ang University Gwang Myeong Hospital, Gyeonggi-do, Republic of Korea.
Purpose: This study aimed to assess the dosimetric advantages of dual arc increments in head and neck volumetric-modulated arc therapy (VMAT) in the Monaco treatment planning system (TPS).
Methods: Three VMAT plans were created for each of the 10 patients by prescribing 70 Gy in 35 fractions with arc increment combinations of 30°/30°, 15°/15°, and 30°/15° in the Monaco TPS. The dose to 0.
Medicine (Baltimore)
October 2024
Department of Radiation Oncology, Sir Run Run Shaw Hospital, Zhejiang University School of Medicine, Hangzhou, China.
This study aimed to investigate the influence of gantry angle increment (Inc) parameters on the Monaco treatment planning system for volumetric modulated arc therapy (VMAT) in right breast cancer, and to determine the optimal Inc parameters. Twenty-three patients with right breast cancer at the Sir Run Run Shaw Hospital from August 2022 to August 2023 were retrospectively selected. Four VMAT plans were generated for each patient incorporating Inc of 10°, 20° (control group), 30°, and 40°, while the other optimization parameters and constraint functions remained constant.
View Article and Find Full Text PDFJ Appl Clin Med Phys
September 2024
Department of Oncology, Daping Hospital, Army Military Medical University, Chongqing, China.
Background: To investigate the effect of the Increment of gantry angle (Inc) parameter setting of the Monaco Treatment planning system (Monaco TPS) on the dosimetry and quality parameters of the volumetric modulated arc therapy (VMAT) program for rectal cancer.
Methods: A retrospective analysis was conducted on 50 patients with rectal cancer who underwent intensity modulated radiation therapy using the Monaco TPS system from 2020 to 2021. Under the same optimization function configuration and other parameter settings, the Inc parameters in the VMAT radiotherapy plan were set to 10°, 20°, 30°, and 40°.
Med Dosim
November 2024
Department of Medical Physics, School of Medicine, University of Patras, 26504, Patras, Greece. Electronic address:
Recently, carbon fiber (CF) has prevailed as the primary material used in radiotherapy couchtops. Modern couchtops incorporate the CF sandwich design, in which 2 thin CF plates sandwich an air-equivalent polymeric foam. Developments in radiotherapy necessitate irradiation from posterior angles through the couchtop.
View Article and Find Full Text PDFRadiol Phys Technol
March 2024
Department of Quantum Medical Technology, Faculty of Health Sciences, Institute of Medical, Pharmaceutical and Health Sciences, Kanazawa University, 5-11-80 Kodatsuno, Kanazawa, Ishikawa, 920-0942, Japan.
Solid-state detectors (SSDs) may be used along with a lead collimator for half-value layer (HVL) measurement using computed tomography (CT) with or without a tin filter. We aimed to compare HVL measurements obtained using three SSDs (AGMS-DM+ , X2 R/F sensor, and Black Piranha) with those obtained using the single-rotation technique with lead apertures (SRTLA). HVL measurements were performed using spiral CT at tube voltages of 70-140 kV without a tin filter and 100-140 kV (Sn 100-140 kV) with a tin filter in increments of 10 kV.
View Article and Find Full Text PDFEnter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!