Objectives: This study aimed to compare the efficacy and safety of the Nit-Occlud PDA device (PFM Medical, Cologne, Germany) to benchmarks designed as objective performance criteria (OPC).

Background: The Nit-Occlud PDA is a nitinol coil-type patent ductus arteriosus (PDA) occluder with a reverse cone configuration, which is implanted using a controlled delivery system.

Methods: Patients with <4-mm minimum diameter PDA were prospectively enrolled in the Pivotal and the Continuing Access Studies from 15 sites in the United States and were followed up for 12 months post-procedure. Investigator-reported outcomes were compared to OPC including a composite success criterion, efficacy criteria of successful closure (clinical and echocardiographic), and safety criteria incidence of adverse events (serious and of total).

Results: The Pivotal Study enrolled patients between November 1, 2002 and October 31, 2005, and the Continuing Access Study enrolled additional patients between September 1, 2006 and October 31, 2007. A total of 357 patients were enrolled, and 347 had successful device implantations. After 12 months, 96.8% had complete echocardiographic closure (OPC = 85%) and 98.1% had clinical closure (OPC = 95%). There were no deaths or serious adverse events (OPC = 1%). The total adverse event rate was 4.7% (OPC = 6%). Composite success was 95.1% in the study patients (OPC = 80%).

Conclusions: Closure of small- and medium-sized PDA with the Nit-Occlud PDA is effective and safe when compared with OPC.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.jcin.2014.06.019DOI Listing

Publication Analysis

Top Keywords

nit-occlud pda
12
pda device
8
patent ductus
8
ductus arteriosus
8
combined multicenter
4
multicenter pivotal
4
pivotal study
4
study continuing
4
continuing access
4
access study
4

Similar Publications

Introduction: Successful closure of patent ductus arteriosus (PDA) can be obtained with surgical ligation or with occlusion via minimally invasive per-catheter techniques. This study was performed to assess feasibility and effectiveness of transjugular PDA occlusion in dogs weighing < 3 kg with a device called Nit-Occlud® PDA.

Animals: Thirteen client-owned dogs.

View Article and Find Full Text PDF

Background: The decision to close patent ductus arteriosus should always be individualized and taken together with the child's family once the risks and benefits of both choices have been exposed.

Objective: This study aims to report the experience and outcomes in patients undergoing endovascular closure of small to medium-size PDA with a Nit-Occlud device in a tertiary referral hospital in Colombia.

Methods: Longitudinal descriptive study, which included all patients under 18 years of age who underwent percutaneous ductal closure with Nit-Occlud device between January 1, 2011, and February 1, 2023.

View Article and Find Full Text PDF

Key Clinical Message: Descending aorta to right atrial (RA) fistula is a rare and distinct clinical entity mimicking patent ductus arteriosus (PDA) and it may lead to rapid development of pulmonary vascular disease. Correct diagnosis and treatment, especially in the presence of other congenital heart defects, is very important. Interventional management is the treatment of choice.

View Article and Find Full Text PDF

Most interventionalists use the Amplatzer Duct Occluder (ADO) or the Nit-Occlud® Coils (NOC) to close patent ductus arteriosus (PDA). Data regarding the success and effect of NOCs in the occlusion of large PDAs are insufficient. We aimed to investigate whether the PDA occlusion of large PDAs using NOC is safe and efficient for all ages.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!