Pulp chamber temperature increase from curing light units: an in vitro study.

J Dent Child (Chic)

Department of Restorative Dentistry, Faculty of Dentistry, Erciyes University, Kayseri, Turkey.

Published: September 2015

Purpose: The purpose of this study was to compare temperature rises in the pulp chamber induced by halogen, plasma arc, and conventional light-emitting diode (LED) curing units with that induced via a new generation LED-curing unit (VALO) in extra power mode.

Methods: A Class I cavity was prepared on the occlusal surface of 80 extracted caries- free mandibular third molars, which were filled with a microhybrid composite. A thermocouple wire was inserted into the pulp chamber of each tooth to measure temperature changes.

Results: The greatest temperature increases were observed during polymerization of composite resin with a halogen curing unit (3.2 degrees Celsius), followed by plasma arc curing (2.07 degrees Celsius) and VALO curing (1.44 degrees Celsius); the lowest temperature rise was with conventional LED curing (1.01 degrees Celsius).

Conclusion: There were no statistically significant differences between conventional LED and VALO curing in extra power mode regarding pulp chamber temperature increases when polymerizing composite resin.

Download full-text PDF

Source

Publication Analysis

Top Keywords

pulp chamber
16
degrees celsius
12
chamber temperature
8
plasma arc
8
led curing
8
extra power
8
temperature increases
8
composite resin
8
valo curing
8
conventional led
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!