Purpose: To examine the effect of misalignment (decentration and tilt) of intraocular lenses (IOLs) on retinal image quality using a water-immersed model eye with corneal spherical aberration adjusted to the values found in normal human eyes (spherical aberration 0.25 μm; pupil diameter 6 mm).
Methods: Three types of IOL holders were prepared. The first was without decentration or tilt, the second had a decentration of 0.5 mm, and the third had a tilt of 5.0°. One spherical IOL and three aspherical IOLs, each with a power of +20 D, were set in the holders and their optical properties (wave front aberration, defocused modulation transfer function, defocused point spread function, and Landolt ring simulations) were compared.
Results: Coma aberrations generated by misaligned IOLs were related to the spherical aberration corrective power of the IOLs. Landolt ring simulations show that the depth of focus increased as spherical aberration increased and that the retinal image quality was degraded by increases in coma aberration.
Conclusion: Coma aberration was generated by IOLs with a large degree of spherical aberration correction, leading to reduced retinal image quality when the IOL was misaligned. This suggests that, in a clinical setting, the quality of vision might be improved by reducing the degree of coma aberration using IOLs that retain, or minimally correct, spherical aberration.
Download full-text PDF |
Source |
---|---|
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC4259511 | PMC |
http://dx.doi.org/10.2147/OPTH.S72053 | DOI Listing |
Ophthalmic Physiol Opt
January 2025
Northeastern University College of Science, Boston, Massachusetts, USA.
Purpose: To assess longitudinal changes in optical quality across the periphery (horizontal meridian, 60°) in young children who are at high (HR) or low risk (LR) of developing myopia, as well as a small subgroup of children who developed myopia over a 3-year time frame.
Methods: Aberrations were measured every 6 months in 92 children with functional emmetropia at baseline. Children were classified into HR or LR based on baseline refractive error and parental myopia.
Photodiagnosis Photodyn Ther
January 2025
Department of Ophthalmology, Shanghai Tenth People's Hospital, School of Medicine, Tongji University Shanghai, 200072, China.. Electronic address:
Purpose: This retrospective study aimed to investigate the subjective and objective visual outcomes following Small Incision Lenticule Extraction (SMILE) surgery in high myopic patients with varying axial lengths (AL).
Methods: The study enrolled 113 highly myopic patients (202 eyes) who underwent SMILE surgery at Shanghai's Tenth People's Hospital from July 2021 to September 2023. Patients were classified into three groups based on the axial length before surgery: Group A (AL < 26mm, 62 eyes), Group B (26mm ≤ AL < 27mm, 88 eyes), and Group C (AL ≥ 27mm, 52 eyes).
J Cataract Refract Surg
January 2025
Vision and Eye Research Institute, School of Medicine, Anglia Ruskin University, Cambridge, UK.
Purpose: To measure visual acuity at three different defocus planes in pseudophakic subjects with varying levels of spherical aberration induced by an adaptive optics visual simulator. The study aimed to simulate Extended Depth of Focus (EDOF) intraocular lenses (IOLs).
Setting: Private hospital (IMO, Barcelona, Spain).
J Cataract Refract Surg
January 2025
University of Plymouth, Plymouth, UK.
Purpose: To evaluate visual outcomes following bilateral implantation of the RayOne EMV intraocular lens with targeted micro-monovision.
Setting: Southend Private Hospital, UK.
Design: Retrospective cohort.
Clin Ophthalmol
January 2025
Department of Ophthalmology, Affiliated Hospital of North Sichuan Medical College, Nanchong, Sichuan Province, 637000, People's Republic of China.
Objective: To assess the visual quality in patients with primary angle-closure glaucoma (PACG) complicated by cataracts after cataract phacoemulsification with intraocular lens (IOL) implantation and goniosynechialysis, and to explore the relationship between pupil size and visual quality.
Methods: A retrospective, non-randomized study was conducted, including 65 PACG patients (75 eyes) who underwent cataract surgery with IOL implantation and goniosynechialysis from July 2021 to June 2023, as well as a control group of cataract-only patients. Visual quality was evaluated using objective and subjective methods at least 3 months postoperatively.
Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!