Analyte quantification with comprehensive two-dimensional gas chromatography: assessment of methods for baseline correction, peak delineation, and matrix effect elimination for real samples.

J Chromatogr A

Environmental Chemistry Modeling Laboratory (LMCE), ENAC, Ecole Polytechnique Fédérale de Lausanne (EPFL), Lausanne, Switzerland; Eawag, Swiss Federal Institute of Aquatic Science and Technology, Dübendorf, Switzerland. Electronic address:

Published: January 2015

Comprehensive two-dimensional gas chromatography (GC×GC) is used widely to separate and measure organic chemicals in complex mixtures. However, approaches to quantify analytes in real, complex samples have not been critically assessed. We quantified 7 PAHs in a certified diesel fuel using GC×GC coupled to flame ionization detector (FID), and we quantified 11 target chlorinated hydrocarbons in a lake water extract using GC×GC with electron capture detector (μECD), further confirmed qualitatively by GC×GC with electron capture negative chemical ionization time-of-flight mass spectrometer (ENCI-TOFMS). Target analyte peak volumes were determined using several existing baseline correction algorithms and peak delineation algorithms. Analyte quantifications were conducted using external standards and also using standard additions, enabling us to diagnose matrix effects. We then applied several chemometric tests to these data. We find that the choice of baseline correction algorithm and peak delineation algorithm strongly influence the reproducibility of analyte signal, error of the calibration offset, proportionality of integrated signal response, and accuracy of quantifications. Additionally, the choice of baseline correction and the peak delineation algorithm are essential for correctly discriminating analyte signal from unresolved complex mixture signal, and this is the chief consideration for controlling matrix effects during quantification. The diagnostic approaches presented here provide guidance for analyte quantification using GC×GC.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.chroma.2014.11.049DOI Listing

Publication Analysis

Top Keywords

baseline correction
16
peak delineation
16
analyte quantification
8
comprehensive two-dimensional
8
two-dimensional gas
8
gas chromatography
8
correction peak
8
gc×gc electron
8
electron capture
8
matrix effects
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!